Caution messages for alarms – Metex 8760CLP Total Free Chlorine & pH Analyzer User Manual

Page 43

background image

AQUAMETRIX INC.

1-800-742-1413 www.aquametrix.com

43

Caution Messages for Alarms

Caution Number

Description

CA7.5

Alarm A, “No Chlorine” alarm

CA7.6

Alarm A, HIGH alarm

CA7.7

Alarm A, LOW alarm

CA7.8

Alarm A, DEVIATION alarm

CA7.9

Alarm A, FAULT alarm

CA8.5

Alarm B, “No Chlorine” alarm

CA8.6

Alarm B, HIGH alarm

CA8.7

Alarm B, LOW alarm

CA8.8

Alarm B, DEVIATION alarm

CA8.9

Alarm B, FAULT alarm

Error Description

Causes

Solutions

Process too caustic for accurate

measurement.

Verify process.

+ Err

pH reading off-scale;

pH > 14.

Large electrode offset.

Service or replace electrode.

Electrode not connected.

Connect electrode or check

connections.

Electrode not responding.

Etch glass electrode. Clean

reference electrode.

- Err

pH reading off-scale;

pH < 0.

Process too acidic to be measured.

Verify process.