KROHNE BM 70 M User Manual

Page 91

background image

BM 70M-EEx/ BM 70Mi-EEx

5

3.1.3 Special

lock

The sealing covers of terminal compartment and electronics compartment are secured by special
locks. The locking screw requires use of an Allen key (4 mm size).

3.1.4 I/O

functions

The following limit values need to be taken into account when connecting the I/O interfaces of the
BM 70 M-EEx level gauges:

BM 70 M-EEx with non-intrinsically safe I/O functions

I/O function

(1)

Nominal values of the non-
certified receiver instrument

Added restriction

Acc. to Standard Installation
and Operating Instructions

Acc. to Standard Installation
and Operating Instructions

Power for receiver
instruments max. 250 V

(1)

Only for connection to circuits with “functional extra-low voltage with safety separation (PELV)“

BM 70 Mi-EEx with intrinsically safe I/O functions

I/O function

Marking

Maximum safety values

Passive
current output

EEx ia IIC or
EEx ib IIC

Ui

30V Ii

250mA

Ci

5nF

Li = negligible

Passive
switching output

EEx ia IIC or
EEx ib IIC

Ui

30V Ii

250mA

Ci

5nF

Li = negligible

Interface
Profibus-PA

(1)

EEx ia IIC or
EEx ib IIC/IIB

Ui

30V Ii

300mA

Ci

5nF

Li = negligible

Interface
Fieldbus FF

(1)

EEx ia IIC or
EEx ib IIC/IIB

Ui

30V Ii

300mA

Ci

5nF

Li = negligible

(1)

suitable for connection to intrinsically safe Fieldbus systems based on the FISCO Model

The intrinsically safe circuits of the I/O functions are galvanically isolated both from ground and
also from each other (test voltage > 500 VAC).
Furthermore, all intrinsically safe circuits are safety-separated from the non-intrinsically safe
power circuits up to a peak voltage of 375 V.

3.2 Flange

systems

3.2.1 Category / Zone allocation

The allowable operating conditions for the flange systems of the BM 70 M-EEx level gauges are
given in the following tables.

Note:

The application restrictions marked with indices should be observed without fail.

Flange system

Allowed operating conditions

Material Version

Antenna

in

Zone

Flange temperature
/ product
temperature [°C]

Max. allowable tank -
pressure
[bar]

0

- 20 ... 60°C

(1)

-0.2 ... 0.1

(1)

(0.8 ... 1.1 bar abs.)

V96 and SA
all certified variants

1

see Sect. 3.3.2

Acc. to Standard Installation
and Operating Instructions

(1)

limited to atmospheric conditions in accordance with EN 50 284