PROFIBUS-PA / FF with current output:

PROFIBUS-PA / FF with switching output:

7.7

Terminating resistor for the RS 485 interface

A selector plug on the RS 485 board allows
optionally a high-resistance termination (pre-
setting) or a terminating resistor of 120 Ohm at
the RS 485 connections A and B. For this
purpose, the housing (flameproof enclosure) of
the BM 70 M signal converter needs to be
opened (see Sect. 9.3) and the selector plug
turned 180°, see sketch on the right.

8 User

interface

8.1 Local

display

In the versions with local display at the BM 70 M signal converter, configuration can be carried
out directly at the device. If a local display is not featured, the PC-CAT program (see Sect. 8.12)
must be used to change the parameter setting (assuming a communicable interface is provided).

The available display is:

Standard display with 2-line LCD display (7 mm high digits for the measured value).

Standard display

1

Numerical display of measured value

2

Alphanumeric display of function/unit

3

6 markers to indicate measuring status

4

3 keys for configuration and error inquiry

5

magnetic sensors for control through the
closed housing
(function same as the 3 keys)

6

Compass field, signals actuation of a key

42

-

Ex-i-supply-

+

+

unit

+

U max.= 30V

current

output

4-20mA

U min.

= 8V

41

-

32 31

Bus-

connection

32 31

switching output

max. 110mA/30V

42 41

Bus-

connection

120

RS485 signal-
output board

high-

impedance

....

....