8.7 Functional

checks

8.7.1 Main menu 2.0 Test functions

Function (Fct.)

Input range

Description

2.0 TEST
2.1 HARDWARE

2.1.1 MASTER

Special function

Test of master hardware.

2.1.2 DISPLAY

Special function

Test of display hardware.

Select

Module DIS

2.1.3

STATUS

Module MW

Display of ID Number and
status bytes.

2.2 CUR.OUTP.I

2.2.1 VALUE I

Display of value

Display of actual value of
the current output.

Options 2

mA

4

mA

6

mA

8

mA

10

mA

12

mA

14

mA

16

mA

18

mA

20

mA

2.2.2 TEST I

22

mA

Output of selected value to
the current output.

NOTE !!! With safety inquiry
because current output is
accessed direct !!!

(Exi = min. 3.6 mA)

2.3 SW.OUTP.

Options OPEN

2.3.1 TEST S

CLOSED

Switches the switching
contact on/off.

NOTE !!! With safety inquiry
because current output is
accessed direct !!!

2.4.0 FIRMWARE

2.4.1 MASTER

Display

Display of master firmware
version.

2.4.2 DISPLAY

Display

Display of master firmware
version.

8.7.2 Hardware

test

Testing of the BM 70 M hardware can be initiated as required with these functions when the
device is operating. When an error is established, a message appears in the display (see Sect.
8.8). This hardware test is carried out automatically every time the BM 70 M is started up.

Fct. 2.1.1 MASTER

Select Function 2.1.1, as described in Sect. 8.2 and 8.3.

"TEST" is shown in the display during the automatic test, followed by "READY" on completion
of the test.

During the test the current output holds the last measured value; digital Communication is not
possible during that time.