8.6.7 Display
Fct. 3.2.1 (or Fct. 1.1.1) FCT.DISP.
This function is used to select the measured variable.
•
LEVEL
•
DISTANCE
•
CONVERSION
•
REFLECTION
Measuring range, see Figure in
Sect. 8.5.
If the conversion value is to be displayed, a conversion table must be programmed first otherwise
a parameter error will occur.
Fct. 3.2.2 (or Fct. 1.1.2) UNIT.LENGTH
Use this function to select the unit for displaying level and distance.
•
m
•
cm
•
mm
•
inch
•
Ft
•
PERCENT
•
BARGRAPH
The "percent display" refers to the scaling range of current output I. Therefore the current output
must also be set to level or distance, otherwise a parameter error will occur.
Under "BARGRAPH", a maximum of 25 segments are displayed in the form of vertical lines in the
first line of the display (0% = 1 segment, 100% = 25 segments).
Fct. 3.2.3 (or Fct. 1.1.3) UNIT.CONV.
Use this function to select the unit for conversion display.
•
m
3
•
litres
•
US Gal
•
GB Gal (Gal = gallons)
•
ft3
•
bbl (petroleum barrels)
•
PERCENT
•
BARGRAPH
•
User Unit (see Fct. 3.2.4)
The "percent display" refers to the scaling range of current output I. Therefore the current output
must also be set to conversion, otherwise a parameter error will occur.
Under "BARGRAPH" in the first line of the display a maximum of 25 segments as vertical lines
are displayed (0% = 1 segment, 100% = 25 segments).
Fct. 3.2.4 USER UNIT
Enter the text for the conversion "user unit" (see Fct. 3.2.3) here. It consists of a maximum of 10
alphanumeric characters.
Characters assignable to each of the 10 places:
•
alpha characters A-Z / a-z
•
numbers 0-9
•
special signs + -
•
blank characters (= underscore)