BM 70 M

Function (Fct.)

Enter range

Default value

Description

Options OFF
ON

3.5.2

EMPTY.SPEC.

RECORD

OFF *

Switch the empty-
tank spectrum (profile
of the empty tank)
on/off, or make new
recording.

3.5.3

TIMECONST.

Value
10... 100 [s]

30 s

Input of the time
constant for
measured-value
filtering (low-pass).

3.5.4

TRACING.VEL. Value

0.01 ... 1.00 [m/Min]

0.10 m/min

Input of the maximum
rate of change in the
level that can occur
in operation.

Options NO

3.5.5

MULT.REFL.

YES

NO

Switch the multi-
reflection identifier
on/off.

Options NO

3.5.6

BD-DETECT

YES

YES

Switch the block
distance (overfill)
detector on/off.

Options OFF

3.5.7 FUNCT. FTB

PARTIAL

OFF

Select function of the
tank bottom tracking
system.

3.5.8 EPSILON R

Enter
1.1000 ... 8.0000

2.0000

Input of the relative
permittivity of the
product for the tank
bottom tracking
system.

Options STORAGE

TANK

3.5.9

TANKTYPE

PROC.TANK.

PROC.TANK.

Select the tank type.

3.6 SW.OUTP.S

Options OFF
LEVEL
DISTANCE
CONVERSION
REFLECTION
ERROR

OPEN

3.6.1 FUNCTION S

ERROR

CLOSED

ERROR
OPEN

Select the function of
the switching output.

Option HIGH

3.6.2 TYPE S

LOW

HIGH

Select type of limit
value for the
switching output.
(not applicable to
Function OFF,
ERROR OPEN or
ERROR CLOSED)