8.8.3 Error messages and their representation in display and at outputs
The following list gives all error messages that can occur:
Error messages
Description of
error
To correct device fault
Output of errors via display and outputs,
depending on setting
Type
Text
Display
Current output I
Switching output S
In 2nd
line of
display
ERROR
MSG.
Fct. 3.2.4
RANGE I
Fct. 3.3.2
FUNCTION S
Fct. 3.6.1
yes
no
4-20/
2=E
4-20/
22=E
4-20 ERROR
OPEN
ERROR
CLOSED
OFF
LEVEL
DISTANCE
CONVERSION
REFLECTION
b) NO
M.VALUE
Evaluable
measuring
signals not
available
If this error is displayed for a
lengthy period, this could mean
application-induced errors.
Check the application conditions!
yes no yes
no yes
no
c) LINE
INTER-
RUPT
Power failure
detected
During restart following a brief
power failure. Display vanishes
as soon as a valid measured
value is again present.
yes no yes
no yes
no
SWEEP
LOW
Restricted
functioning of
microwave
board
Measurement continued,
possibly with reduced accuracy.
If error present for longer time or
"VCO SWEEP" error (see below)
occurs, consult KROHNE
Service!
yes no no
no no
no
WATCH
DOG
Processor
monitoring
If error present for longer time,
this means an error on the CPU
board; needs to be replaced by
KROHNE Service!
yes no no
no no
no
SPECT.
ERR.
Empty
spectrum faulty
The current empty-tank spectrum
does not match the changed
operating parameters. Remedy:
record new empty spectrum (Fct
3.5.2).
yes no no
no no
no
NO
DISP.
Configuration
faulty
Reset parameters for the
displayed units (Fct. 3.2.2,
3.2.3).
yes yes
no
no no
no
CAL.
DATA
EEPROM
error: current
output
calibration data
Calibration data of the current
output need to be checked by
KROHNE Service and reset if
necessary.
yes no yes
(22 mA)
no yes
no
CONFIG.
ERR
EEPROM
error:
parameters
faulty
Check all operating parameters
and reset if necessary. If error
persists: EEPROM defective.
Needs to be replaced by
KROHNE Service!
yes yes
yes
(22 mA)
no yes
yes
d) EEPROM
ERR.
EEPROM
defective
EEPROM defective. Needs to be
replaced by KROHNE Service!
yes no yes
no yes
no
ADC
ERR.
Analog/digital
converter
defective
CPU board defective. Needs to
be replaced by KROHNE
Service!
yes no yes
no yes
no
INTEGR.
Integrator
defective
CPU board defective. Needs to
be replaced by KROHNE
Service!
yes no yes
no yes
no
SIGNAL
DOWN
Measuring
signal too weak
for evaluation
No measuring signal, usually
caused by a defect in the
electronics.
yes no yes
no yes
no
VCO
RISE
Fault on
microwave
board
Microwave board defective.
Needs to be replaced by
KROHNE Service!
yes no yes
no yes
no
VCO
SWEEP
Fault on
microwave
board
Microwave board defective.
Needs to be replaced by
KROHNE Service!
yes no yes
no yes
no
e) FATAL
ERROR
CPU board
defective
After a device start-up. If still
possible, invoke Error List and
note down any other errors.
Consult KROHNE Service!
yes yes
no
no no