8.8.3 Error messages and their representation in display and at outputs

The following list gives all error messages that can occur:

Error messages

Description of
error

To correct device fault

Output of errors via display and outputs,
depending on setting

Type

Text

Display

Current output I

Switching output S

In 2nd
line of
display

ERROR
MSG.
Fct. 3.2.4

RANGE I
Fct. 3.3.2

FUNCTION S
Fct. 3.6.1

yes

no

4-20/
2=E

4-20/
22=E

4-20 ERROR

OPEN
ERROR
CLOSED

OFF
LEVEL
DISTANCE
CONVERSION
REFLECTION

b) NO

M.VALUE

Evaluable
measuring
signals not
available

If this error is displayed for a
lengthy period, this could mean
application-induced errors.
Check the application conditions!

yes no yes

no yes

no

c) LINE

INTER-
RUPT

Power failure
detected

During restart following a brief
power failure. Display vanishes
as soon as a valid measured
value is again present.

yes no yes

no yes

no

SWEEP

LOW

Restricted
functioning of
microwave
board

Measurement continued,
possibly with reduced accuracy.
If error present for longer time or
"VCO SWEEP" error (see below)
occurs, consult KROHNE
Service!

yes no no

no no

no

WATCH

DOG

Processor
monitoring

If error present for longer time,
this means an error on the CPU
board; needs to be replaced by
KROHNE Service!

yes no no

no no

no

SPECT.

ERR.

Empty
spectrum faulty

The current empty-tank spectrum
does not match the changed
operating parameters. Remedy:
record new empty spectrum (Fct
3.5.2).

yes no no

no no

no

NO

DISP.

Configuration
faulty

Reset parameters for the
displayed units (Fct. 3.2.2,
3.2.3).

yes yes

no

no no

no

CAL.

DATA

EEPROM
error: current
output
calibration data

Calibration data of the current
output need to be checked by
KROHNE Service and reset if
necessary.

yes no yes

(22 mA)

no yes

no

CONFIG.

ERR

EEPROM
error:
parameters
faulty

Check all operating parameters
and reset if necessary. If error
persists: EEPROM defective.
Needs to be replaced by
KROHNE Service!

yes yes

yes
(22 mA)

no yes

yes

d) EEPROM

ERR.

EEPROM
defective

EEPROM defective. Needs to be
replaced by KROHNE Service!

yes no yes

no yes

no

ADC

ERR.

Analog/digital
converter
defective

CPU board defective. Needs to
be replaced by KROHNE
Service!

yes no yes

no yes

no

INTEGR.

Integrator
defective

CPU board defective. Needs to
be replaced by KROHNE
Service!

yes no yes

no yes

no

SIGNAL

DOWN

Measuring
signal too weak
for evaluation

No measuring signal, usually
caused by a defect in the
electronics.

yes no yes

no yes

no

VCO

RISE

Fault on
microwave
board

Microwave board defective.
Needs to be replaced by
KROHNE Service!

yes no yes

no yes

no

VCO

SWEEP

Fault on
microwave
board

Microwave board defective.
Needs to be replaced by
KROHNE Service!

yes no yes

no yes

no

e) FATAL

ERROR

CPU board
defective

After a device start-up. If still
possible, invoke Error List and
note down any other errors.
Consult KROHNE Service!

yes yes

no

no no