BM 70 M

35

Function (Fct.)

Enter range

Default value

Description

2.3 SW.OUTPUT

OPEN

2.3.1 TEST S

Options

CLOSED

OPEN

Switching output on/off.

IMPORTANT !!! With
safety inquiry because
of direct access to the
switching output !!!

2.4 FIRMWARE

2.4.1 MASTER

Display

Displays the firmware
version of the master.

2.4.2 DISPLAY

Display

Displays the firmware
version of the display.

3.0 INSTALL.
3.1 BASIS.PARAM

m

cm
mm

inch

Options,
unit

Ft

3.1.1

TANKHEIGHT

Enter
0.50 [m]...max. tank
height

M

10.00 m* /
20.00 m *

To input the tank
height.
The tank height is
defined as the vertical
distance between the
bottom edge of the
flange and the tank
bottom.

The unit entered here is
also used for all other
length inputs.

3.1.2

BLOCKDIST Enter

0.10 [m] ... tank height

0.50 m *

To input the block
distance.
The block distance
marks the non-
measurable range
below the bottom edge
of the flange.

3.1.3 ANTENNA

Options

STANDARD

STANDARD*

Selection of the
antenna type.

3.1.4

ANT.EXTENS Enter

0.00 [m] ... Tank height

0.00 m *

To input the length of
the antenna extension.

3.1.5

DIST.PIECE Enter

0 ... 2000 [mm]

0 mm *

To input the length of
the flange distance
piece in [mm] (high-
temperature version =
120 mm).

Options NO
YES
If "YES":

Enter (restraint)
STILLWELL DIA.

3.1.6

STILLWELL

25 ... 200 [mm]

NO

100 mm

Select: with or without
stilling well. When
"YES" selected, the
inside diameter of the
stilling well is
additionally entered in
[mm].

3.1.7

REF.OFFSET Enter

-10.00 ... + 10.00 [m]

0.00 m

To input the reference
offset (relocation of the
upper reference point).