beautypg.com

KROHNE BM 70 M User Manual

Page 16

background image

16

BM 70 M

Do not position in tank
centreline! (multiple
reflections!)

Do not mount dead centre on
manhole cover! (multiple
reflections!)

Do not position above
internals! (interference
reflections!)

If these recommendations cannot be followed because of the tank geometry, please consult
KROHNE.

B) Stilling well or Wave-Guide

A stilling well or Wave-Guide may be mounted in any position on the tank.
(For design layout: see also Section 6.1.4; for block distance: see Section 3.3).