KROHNE BM 70 M User Manual
Page 37
BM 70 M
37
Function (Fct.)
Enter range
Default value
Description
3.3.4 SCALE 20 mA
Enter
-200.00 ... +200.00
[m]
0.00 ... 99999.99
[m3]
0 ... 100
[Refl.]
10.00 m *
10.00 m3
100
Input of the full-scale
range for output to the
current output.
(Input is dependent on
the selected function
of the current output.
The values for level
and distance include
the tank bottom and
reference offset,
resp.).
Options 1200
Bd.
2400
Bd.
4800
Bd.
9600
Bd.
19200
Bd.
3.3.5
BAUDRATE
38400
Bd.
19200 Bd. *
Select the baud rate
for the Communication
interface (not relevant
to HART
via current
output).
(For Fct. 3.3.7 =
PROFIBUS-PA or
F.FOUND.: deviating
options.)
3.3.6 ADDRESS
Enter
0 ... 255
0
Input of the device
address for
Communication.
(For Fct. 3.3.7 =
PROFIBUS-PA or
F.FOUND.: deviating
input range)
Options HART
®
(MODBUS)
(PROFIBUS-
PA)
3.3.7
PROTOCOL
(F.FOUND.)
HART
®
Select
the
Communications
protocol. (only when
appropriate hardware
provided
3.4 USER DATA
Options GB/USA
D
F
I
E
P
3.4.1
LANGUAGE
S
GB/USA *
Select language to be
used in the display
(English, German,
French, Italian,
Spanish, Portuguese,
Swedish)
Options NO
3.4.2 ENTRY CODE1
YES
NO
Switch the access
lockout for the
configuration menu
on/off.
3.4.3 CODE 1
Enter code
RRREEEUUU
Input of the Entry
Code.
3.4.4 LOCATION
Text
BM70-00
Input of a max. 8-
character device
identifier.
3.5 APPLICAT.
3.5.1 AUTO TANKH.
Special function
Automatic
determination of the
tank height