Technical data – KROHNE OPTIWAVE 6300C EN User Manual
Page 84
8
TECHNICAL DATA
84
OPTIWAVE 6300 C
www.krohne.com
03/2014 - 4000547004 - HB OPTIWAVE 6300 R04 en
Process connections
Thread
G 1½ (ISO 228); 1½ NPT (ASME B1.20.1)
Flange version
Flange version
Flange version
Flange version
EN 1092-1
DN80 in PN40 (Form B1), DN100…200 in PN16 or PN40 (Form B1); others on
request
ASME B16.5
3¨…8¨ in 150 lb RF, 3¨...4¨ in 300 lb RF; others on request
JIS B2220
80…100A in 10K; others on request
Other
Others on request
Electrical connections
Power supply
Terminals output 1
Terminals output 1
Terminals output 1
Terminals output 1 – Non-Ex / Ex i:
Non-Ex / Ex i:
Non-Ex / Ex i:
Non-Ex / Ex i:
14…30 VDC; min./max. value for an output of 22 mA at the terminal
Terminals output 1
Terminals output 1
Terminals output 1
Terminals output 1 – Ex d:
Ex d:
Ex d:
Ex d:
20…36 VDC; min./max. value for an output of 22 mA at the terminal
Terminals output 2
Terminals output 2
Terminals output 2
Terminals output 2 – Non-Ex / Ex i / Ex d:
Non-Ex / Ex i / Ex d:
Non-Ex / Ex i / Ex d:
Non-Ex / Ex i / Ex d:
10…30 VDC; min./max. value for an output of 22 mA at the terminal (additional
power supply needed – output only)
Cable entry
M20×1.5; ½ NPT
G ½ (not for FM- and CSA-approved devices. Not for stainless steel housings.)
M25×1.5 (for stainless steel housings only)
Cable gland
Standard: none
Options: M20×1.5 (for non-Ex and Ex -approved devices with M20×1.5 and M25×1.5
cable entries); others are available on request
Cable entry capacity (terminal)
0.5…1.5 mm²
Input and output
Current output
Current output
Current output
Current output
Output signal
(Output 1)
4…20 mA HART
®
or 3.8…20.5 mA acc. to NAMUR NE 43
4
Output signal
(Output 2 - optional)
4…20 mA (no HART
®
signal) or 3.8…20.5 mA acc. to NAMUR NE 43
Resolution
±3 µA
Temperature drift
Typically 25 ppm/K
Error signal
High: 22 mA; Low: 3.6 mA acc. to NAMUR NE 43