Operation, 5 function description – KROHNE OPTIWAVE 6300C EN User Manual
Page 47
OPERATION
6
47
OPTIWAVE 6300 C
www.krohne.com
03/2014 - 4000547004 - HB OPTIWAVE 6300 R04 en
6.3.5 Function description
A. Quick setup
Menu
No.
Step
Function
Function description
Selection list
Default
A.1 Setup Mode
A.1.1
A.1.1
A.1.1
A.1.1
Complete
Complete
Complete
Complete
This follows the steps given in
This follows the steps given in
This follows the steps given in
This follows the steps given in
the Installation, Empty
the Installation, Empty
the Installation, Empty
the Installation, Empty
Spectrum, Conversion and
Spectrum, Conversion and
Spectrum, Conversion and
Spectrum, Conversion and
Outputs setup modes.
Outputs setup modes.
Outputs setup modes.
Outputs setup modes.
A.1.2
A.1.2
A.1.2
A.1.2
Installation
Installation
Installation
Installation
Follow this procedure to
Follow this procedure to
Follow this procedure to
Follow this procedure to
describe the silo and the
describe the silo and the
describe the silo and the
describe the silo and the
product.
product.
product.
product.
1
Installation Type
The silo material.
Metal / concrete silo, Plastic
silo, Free Air Application
Metal silo
2
Tank Height /
Measuring Range
"Tank Height" is the distance
from the silo connecting flange
face/thread stop to the silo
bottom. "Measuring Range"
(for "Free Air Application" only)
is the maximum distance that a
device must measure.
min-max:
0.20…80 m / 0.66…262 ft
20 m
3
Application Type
The conditions in which the
device is used. If the surface of
the product is flat, select "Flat
surface". If there is a slightly
uneven surface, select
"Medium slope". If the surface
has a steep slope, select
"Steep slope".
Medium slope, Flat surface,
Steep slope
Medium
slope
4
Blocking Distance
A zone given by the user where
it is not possible to measure.
We recommend a blocking
distance at least 300 mm / 12¨
below the antenna.
min-max:
antenna extension (C.1.6) +
50 mm / 2¨…tank height
(C.1.2)
1
2
5
Settings summary
6
Current changes
must be saved or
cancelled before
proceeding
Save, Cancel
3
A.1.3
A.1.3
A.1.3
A.1.3
Empty Spectrum
Empty Spectrum
Empty Spectrum
Empty Spectrum
Fixed and moving objects in the
Fixed and moving objects in the
Fixed and moving objects in the
Fixed and moving objects in the
tank cause interference
tank cause interference
tank cause interference
tank cause interference
signals. Put them through this
signals. Put them through this
signals. Put them through this
signals. Put them through this
filter to correctly measure the
filter to correctly measure the
filter to correctly measure the
filter to correctly measure the
silo contents.
silo contents.
silo contents.
silo contents.
1
Do you have a
completely filled
tank?
If the silo is full, it is not
possible to complete this
procedure. The silo must be
partially filled or empty.
Yes, No
Yes
2
Are all moving
parts in the tank,
e.g. agitators, in
motion?
We recommend that you switch
on moving equipment to filter
all interference signals.
Yes, No
Yes
3
Is your tank
partially filled or
empty?
If the silo is partially filled, the
device must include the silo
contents when it filters the
signal.
Partially Filled, Empty
Partially
Filled