OPERATION
6
55
OPTIWAVE 6300 C
www.krohne.com
03/2014 - 4000547004 - HB OPTIWAVE 6300 R04 en
C.3.4
Output Range
Sets the effective range of output 1
either with or without over-run.
min.-max: 3.8…20.5 mA
(NAMUR), 4…20 mA
4…20 mA
C.3.5
Error Handling
This sets the behaviour of current
output 1 if an error occurs. Hold means
that the output current stays at the
value where the error occurred. Hold is
not available if 3.8…20.5 mA (NAMUR)
is the output range.
3.6 mA, 22 mA, Hold
22 mA
Error Handling
Delay
The time after which the current output
changes to an error value. It shows
that there is a measurement error.
min.-max:
0…900 sec (seconds)
10 sec
C.3.6
HART Address
Any HART
®
address greater than 0 will
activate HART
®
multidrop mode. The
current output stays constant at 4 mA.
min.-max: 0…15
0
C.4 Output 2 (Passive)
C.4.1
Output
Function
Select an output function to scale the
current values. This is not displayed in
the operator mode.
Level, Distance, Volume
(Mass), Ullage Volume (Ullage
Mass), Reflection
Level
C.4.2
4 mA Setting
Give a measurement value to 4 mA.
min.-max:
1
0 m / 0 ft
C.4.3
20 mA Setting
Give a measurement value to 20 mA.
min.-max:
2
Depends on the
output function
C.4.4
Output Range
Sets the effective range of output 2
either with or without over-run.
min.-max: 3.8…20.5 mA
(NAMUR), 4…20 mA
4…20 mA
C.4.5
Error Handling
This sets the behaviour of current
output 2 if an error occurs. Hold means
that the output current stays at the
value where the error occurred. Hold is
not available if 3.8…20.5 mA (NAMUR)
is the output range.
3.6 mA, 22 mA, Hold
22 mA
Error Handling
Delay
The time after which the device shows
there is a measurement error. This
value is set in the output 1 menu.
Read only
Refer to
function C.3.5
C.5 Device Setup
C.5.1
C.5.1
C.5.1
C.5.1
Display
Display
Display
Display
Settings
Settings
Settings
Settings
To display the information you need,
To display the information you need,
To display the information you need,
To display the information you need,
refer to these menu items.
refer to these menu items.
refer to these menu items.
refer to these menu items.
C.5.1.1
Language
Information can be displayed in any of
the 9 languages stored in the device. If
it is necessary to change the display
quickly to English, push [
] for 3
seconds in operator mode.
English, French, German,
Italian, Japanese, Chinese
(simplified), Portuguese,
Russian, Spanish
C.5.1.2
Display Mode
The display screen status will change
after the time given in C.5.1.3. (time
delay). Disable
Disable
Disable
Disable switches off this
functionality, Auto-Off
Auto-Off
Auto-Off
Auto-Off switches off the
display and Default Screen
Default Screen
Default Screen
Default Screen will show
the selected default screen. To set to
the default screen, push [
] for 1
second in the operator mode.
Disable, Auto-Off, Default
Screen
Disable
C.5.1.3
Time Delay
The time after which the display will
switch to the status set in C.5.1.2
(Display Mode).
1, 3, 5, 10 (minutes)
1
Contrast
The contrast control for the display
screen. You can select a shade of gray
between light gray (level 1) and black
(level 9).
Level 1, Level 2, Level 3, Level
4, Level 5, Level 6, Level 7,
Level 8, Level 9
Level 5
Menu
No.
Function
Function description
Selection list
Default