TECHNICAL DATA

8

83

OPTIWAVE 6300 C

www.krohne.com

03/2014 - 4000547004 - HB OPTIWAVE 6300 R04 en

Other conditions

Other conditions

Other conditions

Other conditions

Dielectric constant (εr)

≥1.5

Vibration resistance

IEC 60068-2-6 and EN 50178 (10...57 Hz: 0.075 mm / 57...150 Hz:1g)

Protection category

IP 66/67 equivalent to NEMA type 4X (housing) and type 6P (antenna)

Maximum rate of change

10 m/min / 33 ft/min

Installation conditions

Process connection size

The nominal diameter (DN) should be equal to or larger than the antenna diameter.

If the nominal diameter (DN) is smaller than the antenna, either:

– provide the means to adapt the device to a larger process connection on the silo

(for example, a plate with a slot), or

– use the same process connection, but remove the antenna from the device before

installation and fit it from inside the silo

Process connection position

Make sure that there are not any obstructions directly below the process

connection for the device.

Dimensions and weights

Refer to "Technical data: Dimensions and weights".

Materials

Housing

Standard: Polyester-coated aluminium

Option: Stainless steel (1.4404 / 316L)

1

Wetted parts, including antenna

Horn / Sheet metal horn antenna: Stainless steel (1.4404 / 316L)

Drop antenna: PTFE; PP – a PP or PTFE flange plate protection option is also

available

Process connection

Stainless steel (1.4404 / 316L) – a PP or PTFE flange plate protection option is also

available for the Drop antenna

Gaskets (and o-rings for the

sealed antenna extension option)

PTFE Drop antenna:

PTFE Drop antenna:

PTFE Drop antenna:

PTFE Drop antenna:
FKM/FPM (-40…+150°C / -40…+300°F); Kalrez

®

6375 (-20…+150°C / -4…+300°F);

EPDM (-50°C…+150°C / -58…+300°F)

2

PP Drop antenna:

PP Drop antenna:

PP Drop antenna:

PP Drop antenna:
FKM/FPM (-40…+100°C / -40…+210°F); Kalrez

®

6375 (-20…+100°C / -4…+210°F);

EPDM (-40°C…+100°C / -40…+210°F)

2

Horn / Sheet metal horn antenna:

Horn / Sheet metal horn antenna:

Horn / Sheet metal horn antenna:

Horn / Sheet metal horn antenna:
FKM/FPM (-40…+200°C / -40…+390°F); Kalrez

®

6375 (-20…+200°C / -4…+390°F);

EPDM (-50°C…+150°C / -58…+300°F)

2

Feedthrough

Standard: PEI (-50...+200°C / -58...+390°F – max. range. The feedthrough

temperature limits must agree with the temperature limits of the gasket material

and antenna type. If the distance piece option is not attached, the maximum

temperature is 150°C / 300°F.)

Option: Metaglas

®

(-30...+200°C / -22...+390°F – max. range. The feedthrough

temperature limits must agree with the temperature limits of the gasket material

and antenna type. If the distance piece option is not attached, the maximum

temperature is 150°C / 300°F.)

3

Weather protection (Option)

Stainless steel (1.4301 / 304)