Technical data – KROHNE OPTIWAVE 6300C EN User Manual
Page 83
TECHNICAL DATA
8
83
OPTIWAVE 6300 C
www.krohne.com
03/2014 - 4000547004 - HB OPTIWAVE 6300 R04 en
Other conditions
Other conditions
Other conditions
Other conditions
Dielectric constant (εr)
≥1.5
Vibration resistance
IEC 60068-2-6 and EN 50178 (10...57 Hz: 0.075 mm / 57...150 Hz:1g)
Protection category
IP 66/67 equivalent to NEMA type 4X (housing) and type 6P (antenna)
Maximum rate of change
10 m/min / 33 ft/min
Installation conditions
Process connection size
The nominal diameter (DN) should be equal to or larger than the antenna diameter.
If the nominal diameter (DN) is smaller than the antenna, either:
– provide the means to adapt the device to a larger process connection on the silo
(for example, a plate with a slot), or
– use the same process connection, but remove the antenna from the device before
installation and fit it from inside the silo
Process connection position
Make sure that there are not any obstructions directly below the process
connection for the device.
Dimensions and weights
Refer to "Technical data: Dimensions and weights".
Materials
Housing
Standard: Polyester-coated aluminium
Option: Stainless steel (1.4404 / 316L)
1
Wetted parts, including antenna
Horn / Sheet metal horn antenna: Stainless steel (1.4404 / 316L)
Drop antenna: PTFE; PP – a PP or PTFE flange plate protection option is also
available
Process connection
Stainless steel (1.4404 / 316L) – a PP or PTFE flange plate protection option is also
available for the Drop antenna
Gaskets (and o-rings for the
sealed antenna extension option)
PTFE Drop antenna:
PTFE Drop antenna:
PTFE Drop antenna:
PTFE Drop antenna:
FKM/FPM (-40…+150°C / -40…+300°F); Kalrez
®
6375 (-20…+150°C / -4…+300°F);
EPDM (-50°C…+150°C / -58…+300°F)
2
PP Drop antenna:
PP Drop antenna:
PP Drop antenna:
PP Drop antenna:
FKM/FPM (-40…+100°C / -40…+210°F); Kalrez
®
6375 (-20…+100°C / -4…+210°F);
EPDM (-40°C…+100°C / -40…+210°F)
2
Horn / Sheet metal horn antenna:
Horn / Sheet metal horn antenna:
Horn / Sheet metal horn antenna:
Horn / Sheet metal horn antenna:
FKM/FPM (-40…+200°C / -40…+390°F); Kalrez
®
6375 (-20…+200°C / -4…+390°F);
EPDM (-50°C…+150°C / -58…+300°F)
2
Feedthrough
Standard: PEI (-50...+200°C / -58...+390°F – max. range. The feedthrough
temperature limits must agree with the temperature limits of the gasket material
and antenna type. If the distance piece option is not attached, the maximum
temperature is 150°C / 300°F.)
Option: Metaglas
®
(-30...+200°C / -22...+390°F – max. range. The feedthrough
temperature limits must agree with the temperature limits of the gasket material
and antenna type. If the distance piece option is not attached, the maximum
temperature is 150°C / 300°F.)
3
Weather protection (Option)
Stainless steel (1.4301 / 304)