24

Automatic Measurement (Delay)

Background

Using the automatic measurement function, you
can easily measure the waveform, including the
time distance between two waveforms.

Connection

CH1

CH2

CH1/CH2 Scale

50ns/div, 200mV/div
(if CH1 is still unstable, increase the trigger level)

1. Press the Measurement key to show the Measurement menu.

Measure

2. Press F1 → F3 → F3 (Delay FRR) to select the measurement item.

Press F5 (Previous menu) to go back and observe the delay time.

Vpp

1:344mV
2:24.0mV

F 1

Time

Frequency

F 3

Delay

DelayFRR

F 3

Previous

Menu

F 5

FRR

3. Press F2 → F3 → F3 and use VARIABLE knob to select Delay

LRR. Press F5 (Previous menu) to go back and observe the delay.

Delay

DelayFRR

F 3

Previous

Menu

F 5

F 3

Delay

DelayLRR

F 3

F 2

Vavg

1:99.3mV
2:4.28mV

VARIABLE

LRR

Demonstration completed