GDS-2000 Series Training Manual

22

1GSa/s & Single Shot

Background

By using a very fast sampling rate combined with
single shot trigger mode, you can accurately
reconstruct weak signals which are buried in
larger waveform. Here you can observe a tiny
pulse sitting on top of PECL, 3.3V DC signal.

Connection

CH1

(BNC connector on the demo board)

This demo uses a BNC to BNC cable instead of probe.

CH1 Scale

25ns/div, 50mV/div (1GSa/s)

1. Press the CH2 key twice and turn it OFF.

CH2

CH2

2. Press the CH1 key → F1 (Coupling) and select AC (~). A pulse

signal becomes visible on the display.

CH1

Coupling

F 1

Adjust the CH1 vertical position
if necessary.

POSITION

3. Press the Trigger MENU key → F3 (Mode) repeatedly and select

Single trigger mode.

Mode

Single

F 3

MENU

4. Press the Run/Stop key to initiate a single trigger. A pulse signal

becomes visible in a clear manner.