Use and Wear
CAUTION
• Do not exceed +30 dBm into the RF INPUT or +30 dBm reverse power into the
TG OUTPUT.
• Do not place any heavy object on the instrument.
• Avoid severe impacts or rough handling that could damage the GSP-810.
• Use electrostatic discharge precautions while handling and making
connections to the GSP-810.
• Do not place wires into the connectors of the GSP-810, only mating connectors
and adapters.
• Do not block or obstruct cooling fan vent opening on side panels or on the rear
panel of unit.
1) Disassembly of the Instrument
• Do not disassemble the instrument; refer the instrument to a factory approved
service facility only.
2) AC Power Input
CAUTION
• AC input should be within the range of selected line voltage +/- 10%.
• Insure the correct fuse is installed prior to applying voltage for the first time -
90 V ~ 132 VAC input : T 1A / 250V
198 ~ 250 VAC input : T0.5A / 250V
• Check the line voltage setting on the rear panel. If the line voltage does not
match input voltage, change as follows:
a)
Remove AC Power Cord;
b)
Open cover of AC socket with flat blade screwdriver;
c)
Remove selector Cam Drum and rotate to the correct voltage selection
d)
Replace Cam Drum.