GSP-810 User Manual

21

SETUP SCREEN

The setup screen provides access to various parameters. The setup screen

functions are accessed by pressing the SHIFT RBW key. The trace will stop

updating while in the setup screen. The parameters are accessible using the

SPINNER. The following identifies the various parameters:

1. CAL Signal

The “CAL Signal” provides access to an internal amplitude calibrator signal

of 80MHz. Pressing the

Î key toggles the reference signal between “Off”

and “On”.

2. Option Report

The “Opt Report” provides access to the current configuration of the GSP-

810 including optional hardware that is installed. Pressing the

Î key

displays a screen showing Software Version (SW), Firmware Configuration

(FW), Tracking Generator Configuration (TG, where “N” means option not

installed), Receiver (R), Power Meter (PM) and Reference Signal (REF).

The “TG”, "R" and "PM" labels will be followed by “N” if the option is not

installed. The “REF” label is followed by two fields that identify the status of

a 10 MHz system frequency reference signal that can be applied to the unit.

The first field will display “STD” for standard and “IMP” for improved. The

second field will display “INT” for internal 10MHz reference, or “EXT” for

external 10MHz reference.

3. Test All

The “Test all” field automatically checks the LO statuses of LO1, LO2, and

LO3. If all three LO’s are locked, the “Locked” is displayed. If one of the

three is unlocked, then “Unlocked” is displayed.

4. LO1 Status

This displays the status of LO1. An internal status is monitored to determine

if it is “Locked” or “Unlocked” with the result being displayed.