SETUP SCREEN
The setup screen provides access to various parameters. The setup screen
functions are accessed by pressing the SHIFT RBW key. The trace will stop
updating while in the setup screen. The parameters are accessible using the
SPINNER. The following identifies the various parameters:
1. CAL Signal
The “CAL Signal” provides access to an internal amplitude calibrator signal
of 80MHz. Pressing the
Î key toggles the reference signal between “Off”
and “On”.
2. Option Report
The “Opt Report” provides access to the current configuration of the GSP-
810 including optional hardware that is installed. Pressing the
Î key
displays a screen showing Software Version (SW), Firmware Configuration
(FW), Tracking Generator Configuration (TG, where “N” means option not
installed), Receiver (R), Power Meter (PM) and Reference Signal (REF).
The “TG”, "R" and "PM" labels will be followed by “N” if the option is not
installed. The “REF” label is followed by two fields that identify the status of
a 10 MHz system frequency reference signal that can be applied to the unit.
The first field will display “STD” for standard and “IMP” for improved. The
second field will display “INT” for internal 10MHz reference, or “EXT” for
external 10MHz reference.
3. Test All
The “Test all” field automatically checks the LO statuses of LO1, LO2, and
LO3. If all three LO’s are locked, the “Locked” is displayed. If one of the
three is unlocked, then “Unlocked” is displayed.
4. LO1 Status
This displays the status of LO1. An internal status is monitored to determine
if it is “Locked” or “Unlocked” with the result being displayed.