Specifications – GW Instek GOS-630FC User Manual
Page 19
APPENDIX
37
4. If necessary, replace the fuse that is attached to the fuse holder.
115
230
5. Push the fuse holder back into the socket.
230
115
Fuse rating AC 115V: T0.63A/250V, AC 230V: T0.315A/250V
GOS-630FC
User
Manual
38
Specifications
The specifications apply when the GOS‐630FC is powered on for at
least 30 minutes under +20°C – +30°C.
Vertical
Sensitivity
accuracy
1mV/DIV to 2mV/DIV: ±5%
5mV/DIV to 5V/DIV: ±3%
Bandwidth
1mV/DIV to 2mV/DIV: DC to 7MHz
5mV/DIV to 5V/DIV: DC to 30MHz
AC coupling
> 10Hz (reference: 100kHz, 8DIV, –3dB)
Rise time
1mV/DIV to 2mV/DIV: Approx. 50nS
5mV/DIV to 5V/DIV: Approx. 11.7nS
Input impedance Approx. 1M ohm // 25pF.
Square
wave
characteristics
≤ 5% Overshoot at 10mV/DIV
Other ranges: 5% added to the above
Linearity
±
0.1DIV when moving 2 DIV at center
Vertical mode
CH1, CH2, DUAL, ADD
Chop frequency Approx. 250kHz
Input coupling
AC, GND, DC
Max input voltage CAT II 300Vpeak (AC: ≤ 1kHz)
Max
effective
readout
Probe1:1 40Vpp (14Vrms Sine wave)
Probe10:1 400Vpp (140Vrms Sine wave)
Common mode
rejection ratio
≥ 50:1 at 50kHz sine wave
(CH1 and CH2 vertical scales are equal)
Channel
isolation
@ 5mV/DIV
>1000:1 at 50kHz
>30:1 at 30MHz
CH1 signal output ≥ 20 mV/DIV @ 50Ω, 50Hz to 5MHz
CH2 INV BAL.
≤
1 DIV (Reference at center graticule)
Trigger
Trigger source
CH1, CH2, LINE, EXT
Coupling
AC: 20Hz to full bandwidth
Slope
+ / -
Sensitivity
20Hz to 2MHz : 0.5 DIV
(TRIG-ALT:2 DIV, EXT: 200mV)
2MHz to 30MHz: 1.5 DIV
(TRIG-ALT:3 DIV, EXT: 800mV)
TV: Sync pulse > 1 DIV (EXT: 1V)
Triggering modes AUTO, NORM, TV-V, TV-H
EXT TRIG input
Input impedance: Approx.:1MΩ // 25pF
Max. input: CAT II 300Vpeak < 1kHz