MEASUREMENT

33

5. To adjust the Y‐axis position and 

deflection, use the CH2 vertical 
POSITION knob (position) and 
CH2 VOLTS/DIV knob 
(deflection). 

Waveform Magnification

Background

The oscilloscope can magnify the waveform by ten 
times in the horizontal direction. The 
magnification is useful for observing complex 
signals. 

Steps

1. Make sure the waveform is shown on the 

display. 

2. Press the ×10 MAG switch. The 

waveform becomes magnified 
by ten times in the horizontal 
direction. 

x10 magnification

Any part can be covered by using the POSITION control

GOS-630FC

User

Manual

34

F

AQ

The probe waveform is distorted.

You might need to compensate the probe. For details, see page26. 
Note that frequency accuracy and duty factor are not specified for 
the probe compensation waveform and therefore it should not be 
used for reference purpose. 

The trace line does not appear on the display.

Make sure that the trigger mode is in the AUTO mode. In the 
NORMAL mode, the trace does not appear unless a trigger 
condition occurs. 

The alternate trigger (TRIG ALT switch) does not work.

Make sure that the ALT/CHOP switch is released (ALT position). 
The TRIG ALT switch does not work in the CHOP mode. 

The frequency counter in the LCD display does not work.

Make sure that the TRIG ALT switch is not pressed. The frequency 
counter does not work in the alternate trigger mode. 

The TV trigger does not work.

Make sure that the video synchronization signal is positive. The TV‐
V/TV‐H trigger works only when the synchronization signal is 
negative.