SETUP

25

CH1/CH2
VARIABLE
Knob

CAL 

CH1/CH2
Coupling
Switch

AC 

Trigger
SOURCE Switch

CH1 

Trigger SLOPE
Switch

+ (Positive slope) 

TRIG ALT
Switch

Released (alternating trigger 
disabled) 

Trigger MODE
Switch

AUTO 

TIME/DIV
Knob

0.5ms/DIV 

Horizontal
SWP.VAR Knob

CAL 

Horizontal
POSITION
Knob

Center of the display 

x10 MAG
Switch

Released (x10 maginification 
disabled) 

GOS-630FC

User

Manual

26

Set up & Probe Compensation

Follow these steps and properly set up the oscilloscope, make sure it 
is functional, and compensate the probe. 

1 Power

on Press the Power switch and turn 

on the oscilloscope’s power. The 
LED lights when the power is on. 
The trace line starts to appear after 
20 – 30 seconds. 

2 Intensity

and focus
adjustment

Use the INTEN knob and FOCUS 
knob to adjust the trace line 
appearance to the best condition. 

3 Probe

connection

Connect the probe between the 
CAL output and the CH1 input 
terminal. Select x10 (attenuation) 
on the probe. 

Select x10 (attenuation) on the probe. 

4 Probe

attenuation

x1

0

x1

5 Coupling

mode
selection

Select the AC position in the CH1 
AC‐DC‐GND switch. A square 
waveform appears in the display.