Meas channel > moving average – Yokogawa Removable Chassis DX1000N User Manual

Page 68

background image

66

IM 04L41B01-02E

Setup Item

Selectable Range or Selections

Default Value

Setting

Mode=DI

Range
Span Lower
Span Upper

Level/Cont
0, 1
0, 1

Level
0
1

Mode=1-5V

Range
Span Lower
Span Upper
Scale Lower
Scale Upper
Unit
Low-cut

1-5V
0.800 to 5.200
0.800 to 5.200
–30000 to 30000, decimal position :0 to 4
–30000 to 30000, decimal position :0 to 4
6 characters or less
On/Off

1-5V
1.000
5.000
0.00
200.00

Off

Mode=Sqrt

Range
Span Lower
Span Upper
Scale Lower
Scale Upper
Unit
Low-cut
Low-cut value

20mV/60mV/200mV/2V/6V/20V/50V
Depends on the range.
Depends on the range.
–30000 to 30000, decimal position: 0 to 4
–30000 to 30000, decimal position: 0 to 4
6 characters or less
On/Off
0.0 to 5.0



0.00
200.00

Off
0.5

Alarm

1, 2, 3, 4
Type

Value
Relay
Number

Detect

On/Off
H:High/L:Low/h:delta H/l:delta L/R:rate H/

r:rate L/T:delayH/t:delayL
Numerical value
On/Off
I01/.../I06 (Depends on the option.)
S01/S02/S03/.../S29/S30
On/Off

Off
H


Off
I01

On

Meas channel > Tag, Memory, Delay

Setup Item

Selectable Range or Selections

Default Value

Setting

Without /AS1

First-CH, Last-CH

Channel number

001

Tag > Comment

*2

32 characters or less

*2

Tag > No.

*2

16 characters or less

Memory sample > On/Off

On/Off

On

Alarm delay > Time

1 to 3600

10

With /AS1

*3

The First-CH, Last-CH, Tag, and Memory sample items are the same as on a DX without the /AS1 option.
Alarm delay > Time

1 to 3600 when the unit is seconds. 1 to 24 when

the unit is hours.

10

Alarm delay > Unit

Hour/Sec

Sec

Meas channel > Moving average

Setup Item

Selectable Range or Selections

Default Value

Setting

First-CH, Last-CH

Channel number

001

Moving average > On/Off

On/Off

Off

Moving average > Count

2 to 400

2

Setup Items and Default Values