beautypg.com

Wiring procedure, Arrangement of the input terminals – Yokogawa Removable Chassis DX1000N User Manual

Page 55

background image

3

IM 04L41B01-02E

Installation and W

iring

Recommended Replacement

Periods for W

orn Parts

Setup Items and

Default V

alues

Wiring Procedure

A terminal cover is screwed in place on the measuring input terminal block on the rear
panel. A label indicating the terminal arrangement is affixed to the cover.

1.

Turn OFF the DX and remove the terminal cover.

2.

Connect the signal wires to the terminals.

R

ecommended torque for

tightening the screws

Screw terminals

0.9 to 1.0 N•m

Clampted terminals 0.22 to 0.25 N•m

3.

Replace the terminal cover and fasten it with screws. The proper torque for

tightening the screws is 0.6 N•m.

Note

Input signal wires of diameter less than or equal to 0.3 mm may not be secured firmly for
clamped terminals (/H2). Fold over the conducting section of the wire, for example, to make
sure that the wire is securely connected to the clamped terminal.

Arrangement of the Input Terminals

Input terminal block

+/A

/b

CH1

CH2

CH3

CH4

CH1

CH2

CH3

CH4

CH5

CH6

-

/B

+/A

/b

-

/B

CH1

CH2

CH5 CH3 CH1

CH6 CH4 CH2

CH3 CH1

CH4 CH2

/b

+/A

-

/B

/b

+/A

-

/B

/b

+/A

-

/B

+/A

/b

CH1

CH2

-

/B

Screw terminal

Input terminal block of

the DX1002 and DX1002N

Input terminal block of

the DX1004 and DX1004N

Input terminal block of

the DX1006 and DX1006N

Input terminal block of

the DX1012 and DX1012N

Clamp terminal

CH1

CH2

CH3

CH4

CH5

CH6

CH7

CH8

CH9

CH11

CH10

CH12

+/A

/b

-

/B

CH5 CH3 CH1

CH6 CH4 CH2

CH9 CH7

CH10

CH11

CH12

CH8

/b

+/A

-

/B

Installation and Wiring

This manual is related to the following products: