beautypg.com

4 iecex certification, Iecex certification -13 – Yokogawa FVX110 User Manual

Page 23

background image

<2. Handling Cautions>

2-13

IM 01S01C01-01EN

(6) Name Plate

MODEL
SUFFIX

SUPPLY

OUTPUT

STYLE

NO.

F0213.ai

● Name plate

● Tag plate for flameproof type

● Tag plate for intrinsically safe type

AFTER DE-ENERGIZING, DELAY 5 MINUTES

BEFORE OPENING.

WHEN THE AMBIENT TEMP. ≥ 65ºC,

USE THE HEAT-RESISTING CABLES ≥ 90ºC

No. KEMA 10ATEX0157

Ex d IIC T6, Ex tD A21, IP6X T80ºC

Enlcosure: IP66, IP67

Tamb.(GAS) -50(-15) to 75 ºC

Tamb.(Dust) -40(-15) to 75 ºC

No. DEKRA 11ATEX0022 X

Ex ia IIB/IIC T4 Ga

Ex ia IIIC T80ºC Da IP6X

Tamb: -40 to 60ºC

ENCLOSURE: IP66/IP67

FISCO Field device

Entity Parameters Ui=24V, Ii=250mA, Pi=1.2W, Ci=3.52nF, Li=0µH

MODEL: Specified model code.

STYLE: Style code.

SUFFIX: Specified suffix code.

SUPPLY: Supply voltage.

OUTPUT: Output signal.

NO.: Serial number and year of production

*1

.

TOKYO 180-8750 JAPAN:

The manufacturer name and the address

*2

.

*1: The first digit in the final three numbers of the serial

number appearing after “NO.” on the name plate

indicates the year of production. The following is an

example of a serial number for a product that was

produced in 2010:

The year 2010

91K819857 032

*2: “180-8750” is the Zip code for the following address.

2-9-32 Nakacho, Musashino-shi, Tokyo Japan

2.8.4 IECEx Certification

a. IECEx Flameproof Type

Caution for IECEx flameproof type.

Note 1. FVX110 Fieldbus Segment Indicator with

optional code /SF25 are applicable for use

in hazardous locations:

• No. IECEx KEM10.0071

• Applicable Standard: IEC60079-0(:2004),

IEC60079-1(:2003)

• Type of Protection and Marking Code:

Ex d IIC T6

• Temperature Class: T6

• Enclosure: IP66 and IP67

• Ambient Temperature for gas-proof:

–50* to 75ºC (T6)

* –15ºC when O-ring material is Fluoro-rubber.

Note 2. Wiring

• In hazardous locations, the cable entry

devices shall be of a certified flameproof

type, suitable for the conditions of use and

correctly installed.

• Unused apertures shall be closed with

suitable flameproof certified blanking

elements.

Note 3. Operation

• WARNING:

AFTER DE-ENERGIZING, DELAY 5

MINUTES BEFORE OPENING.

• WARNING:

WHEN AMBIENT TEMPERATURE ≥ 65ºC,

USE THE HEAT-RESISTING CABLES ≥

90ºC.

• Take care not to generate mechanical

sparking when accessing to the instrument

and peripheral devices in a hazardous

location.

Note 4. Maintenance and Repair

• The instrument modification or parts

replacement by other than authorized

representative of Yokogawa Electric

Corporation is prohibited and will void IECEx

Certification.