4 iecex certification, Iecex certification -13 – Yokogawa FVX110 User Manual
Page 23

<2. Handling Cautions>
2-13
IM 01S01C01-01EN
(6) Name Plate
MODEL
SUFFIX
SUPPLY
OUTPUT
STYLE
NO.
F0213.ai
● Name plate
● Tag plate for flameproof type
● Tag plate for intrinsically safe type
AFTER DE-ENERGIZING, DELAY 5 MINUTES
BEFORE OPENING.
WHEN THE AMBIENT TEMP. ≥ 65ºC,
USE THE HEAT-RESISTING CABLES ≥ 90ºC
No. KEMA 10ATEX0157
Ex d IIC T6, Ex tD A21, IP6X T80ºC
Enlcosure: IP66, IP67
Tamb.(GAS) -50(-15) to 75 ºC
Tamb.(Dust) -40(-15) to 75 ºC
No. DEKRA 11ATEX0022 X
Ex ia IIB/IIC T4 Ga
Ex ia IIIC T80ºC Da IP6X
Tamb: -40 to 60ºC
ENCLOSURE: IP66/IP67
FISCO Field device
Entity Parameters Ui=24V, Ii=250mA, Pi=1.2W, Ci=3.52nF, Li=0µH
MODEL: Specified model code.
STYLE: Style code.
SUFFIX: Specified suffix code.
SUPPLY: Supply voltage.
OUTPUT: Output signal.
NO.: Serial number and year of production
*1
.
TOKYO 180-8750 JAPAN:
The manufacturer name and the address
*2
.
*1: The first digit in the final three numbers of the serial
number appearing after “NO.” on the name plate
indicates the year of production. The following is an
example of a serial number for a product that was
produced in 2010:
The year 2010
91K819857 032
*2: “180-8750” is the Zip code for the following address.
2-9-32 Nakacho, Musashino-shi, Tokyo Japan
2.8.4 IECEx Certification
a. IECEx Flameproof Type
Caution for IECEx flameproof type.
Note 1. FVX110 Fieldbus Segment Indicator with
optional code /SF25 are applicable for use
in hazardous locations:
• No. IECEx KEM10.0071
• Applicable Standard: IEC60079-0(:2004),
IEC60079-1(:2003)
• Type of Protection and Marking Code:
Ex d IIC T6
• Temperature Class: T6
• Enclosure: IP66 and IP67
• Ambient Temperature for gas-proof:
–50* to 75ºC (T6)
* –15ºC when O-ring material is Fluoro-rubber.
Note 2. Wiring
• In hazardous locations, the cable entry
devices shall be of a certified flameproof
type, suitable for the conditions of use and
correctly installed.
• Unused apertures shall be closed with
suitable flameproof certified blanking
elements.
Note 3. Operation
• WARNING:
AFTER DE-ENERGIZING, DELAY 5
MINUTES BEFORE OPENING.
• WARNING:
WHEN AMBIENT TEMPERATURE ≥ 65ºC,
USE THE HEAT-RESISTING CABLES ≥
90ºC.
• Take care not to generate mechanical
sparking when accessing to the instrument
and peripheral devices in a hazardous
location.
Note 4. Maintenance and Repair
• The instrument modification or parts
replacement by other than authorized
representative of Yokogawa Electric
Corporation is prohibited and will void IECEx
Certification.