Controls and indicators on the hmia, Rs232, T1 out – ZOLL Thermogard XP IVTM System Operation Manual User Manual
Page 99: System
Accesories
-
HMIA
5
The set-up is now complete.
Controls and indicators on the HMIA
The HMIA has the following control/indicators: a reset switch, a green power
indicator and a red warning light.
The front panel of the HMIA is illustrated in the figure below.
Figure 3
Front Panel of the HMIA
The meaning of the symbols on the front panel are as described in the table
below
Table 1 HMIA Control/Indicators
Symbol Description
Function
RST
Reset Button
This is a reset button. Lightly push the tip of a
paperclip, or similar sized device, into the hole
until you feel a “click” and then withdraw the
paper clip.
Caution LED
This red light indicates that there is a
malfunction in the HMIA.
Power LED
This green light indicates that the HMIA is
receiving power from the System.
RS232
RS-232 Serial port connector
cable
This short extension cable is connected to the
RS-232 data port connector on the back of the
display head of the System.
T1 out
Patient temperature output
connector
This connector accepts the custom Alsius
interface cable.
SYSTEM
System umbilical cord
connections
One connector is attached to the round, multi-
pin connector on the back of the display head
of the System via a short cable. The other
connector mates with the umbilical cord of the
System.
600479-001 Rev 2 99/147