Temperature jumps between 26, C and 49, C on hospital monitor – ZOLL Thermogard XP IVTM System Operation Manual User Manual
Page 107: Hirose connector does not fit

Accesories
-
HMIA
13
6. If it passes the test, then press the reset button (
RST
).
7. Wait for 20 seconds and then repeat step 3 above.
8. If the resistance value is still wrong, then disconnect the RS232 cable
from the rear of the display housing, wait for 15 seconds and then
reconnect.
9. Wait for an additional 20 seconds for the HMIA to re-establish
communication with the System
10. If the resistance value is wrong then return the HMIA to ZOLL for service.
Temperature jumps between 26
O
C and 49
O
C On Hospital
Monitor
There has been an internal component failue. Return the HMIA to ZOLL for
service.
Hirose Connector does not fit
There are two versions of System Hirose connectors in clinical use: a 6-pin and
an 8-pin. If the Hirose connector does not fit, count the number of pins in the
HMIA cable. If this is not the same as the number of pins in the umbilical cable
of the System you have the wrong version of HMIA. If you have the wrong
version of the HMIA, then contact ZOLL to arrange for a replacement.
Table 3 HMIA Part Numbers
601033–001 8-pin circular Hirose connector
601033–002 6-pin circular Hirose connector
600479-001 Rev 2 107/147