ZOLL Thermogard XP IVTM System Operation Manual User Manual
Page 147
Specifications
3
SALINE COOLANT CIRCUIT
Priming volume:
200 ml
Heat exchanger:
Disposable stainless steel coil
Priming source:
Sterile saline solution (hospital-provided)
Patient connection:
Directional Luer connections on
72 in. (183 cm) lines
Pump tubing:
Roller pump compatible with directional fittings
Sterility:
Gamma sterilized
Saline alarm:
Reservoir level detection & alarm system
Coolant circuit operating life:
Replace disposable components after seven (7)
days of continuous use
EQUIPMENT CLASSIFICATIONS
Type of protection against moisture:
Ordinary
Type of protection against
electric shock:
Type BF for temperature inputs
Type B for catheter connections
Protection class:
1
Mode of operation:
Continuous
APPROVED PATIENT TEMPERATURE
PROBES
Temperature Probe Standard
YSI-400
YSI-400 Temperature probes verified by
ZOLL to work with the System
MALINCKRODT MEDICAL FOLEY PROBE 8F (200818-001)
MALINCKRODT MEDICAL FOLEY PROBE 10F (200818-002)
MALINCKRODT MEDICAL FOLEY PROBE 12F (200818-003)
MALINCKRODT MEDICAL FOLEY PROBE 14F (200818-004)
MALINCKRODT MEDICAL FOLEY PROBE 16F (200818-005)
MALINCKRODT MEDICAL FOLEY PROBE 18F (200818-006)
MALINCKRODT MEDICAL GP/RECTAL PROBE 9F (200820-
001)
SMITH MEDICAL FOLEY PROBE 10 FR (600224-001)
SMITH MEDICAL FOLEY PROBE 12 FR (600224-002)
SMITH MEDICAL FOLEY PROBE 14 FR (600224-003)
SMITH MEDICAL FOLEY PROBE 16 FR (600224-004)
SMITH MEDICAL FOLEY PROBE 18 FR (600224-005)
SMITH MEDICAL GP/RECTAL PROBE 9 FR (600225-001)
SMITH MEDICAL GP/RECTAL PROBE 12 FR (600225-002)
YSI-400 Temperature probes verified by
ZOLL to work with the System and Rüsch
patient interconnection cable (Rüsch P/N
179516).
Rüsch Sensor Temperature Foley Catheters (CH 8, 10, 12, 14,
16, 18) – Rüsch P/N 179360.
Rüsch Temperature Supracath (CH10, 12, 14) – Rüsch P/N
179370.
ALSIUS START UP KIT
Tubing set including heat exchange coil,
airtrap and roller pump tubing:
Model CG-500D
600479-001 Rev 2 147/147