ZOLL Thermogard XP IVTM System Operation Manual User Manual
Page 17
Introduction
5
1. Display
Screen
2. Alarm Indicator LED.
3. Mute
Button
4. Power On Indicator LED.
5. Target
Temp
Button.
6. Standby / Run Button.
7. Rate Deg / HR Button.
8. Press for Menu / Enter Knob
Figure 2-2. Controls and Display.
Power Indicators
An indicator lamp on the control panel is illuminated when power is switched on.
A second power-on indicator is mounted directly above the power switch on the
rear of the console.
Alarm Indicators
The System typically notifies users of alarm conditions in two ways. When an
alarm occurs, the screen displays an alarm message, and an alarm annunciator
produces an audible alarm tone (beep). The alarm tone can be temporarily
muted by the user, but it cannot be turned off.
If the nature of the failure prevents the System from displaying an alarm
message, the alarm indicator on the control panel will be illuminated.
Control Buttons
The display head features four pushbuttons that are used to control System
functions. To provide confirmation, each time a button is pressed, a “key click”
sound is produced by the annunciator.
Target Temp
Press the “Target Temp” button to display a screen that allows you to set the
patient’s target temperature. You may set a target temperature between 31º C
and 38º C (87.8º F and 100.4º F).
Rate Deg/Hr
Press the “Rate Deg/Hr” button to display a screen that allows you to set the
cooling/warming rate (expressed in degrees per hour). You may set a cooling/
warming rate between 0.10º C/hr and 0.65º C/hr (0.18º F/hr and 1.17º F/hr).
Standby/Run