DA1855A Differential Amplifier
82
922258-00 Rev A
Certifications
This section certifies the instrument’s Electromagnetic Compatibility (EMC), Safety and
Environmental compliances.
EMC Compliance
EC
D
ECLARATION
O
F
C
ONFORMITY
-
EMC
The amplifier meets intent of EC Directive 2004/108/EC for Electromagnetic Compatibility.
Compliance was demonstrated to the following specifications as listed in the Official Journal of the
European Communities:
EN 61326-1:2006, EN 61326-2-1:2006 EMC requirements for electrical equipment for measurement,
control, and laboratory use.
1
Electromagnetic Emissions:
CISPR 11:2003, Radiated and Conducted Emissions Group 1, Class A
2 3
Electromagnetic Immunity:
EN 61000-4-2:2001 Electrostatic Discharge, 4 kV contact, 8 kV air, 4 kV vertical/horizontal coupling
planes
4
EN 61000-4-3:2006 RF Radiated Electromagnetic Field, 3 V/m, 80-1000 MHz; 3 V/m, 1400 MHz - 2
GHz; 1 V/m, 2 GHz - 2.7 GHz
4
EN 61000-4-4:2004 Electrical Fast Transient/Burst, 1 kV on power supply lines, 0.5 kV on I/O signal
data and control lines
4
EN 61000-4-5:2006 Power line Surge, 1 kV AC Mains, L-N, L-PE, N-PE
4
EN 61000-4-6:2007 RF Conducted Electromagnetic Field, 3 Vrms, 0.15 MHz - 80 MHz
4
EN 61000-4-11:2004 Mains Dips and Interruptions, 0%/1 cycle, 70%/25 cycles, 0%/250 cycles
4 5
1 To ensure compliance with all applicable EMC standards, use high quality shielded interface cables.
2 Emissions which exceed the levels required by this standard may occur when the amplifier is connected to a
test object.
3 This product is intended for use in nonresidential areas only. Use in residential areas may cause
electromagnetic interference.
4 Meets Performance Criteria “B” limits of the respective standard: during the disturbance, product undergoes
a temporary degradation or loss of function or performance which is self-recoverable.
5 Performance Criteria “C” applied for 70%/25 cycle voltage dips and for 0%/250 cycle voltage interruption
test levels per EN61000-4-11.