Input-Resistance

PRx:PINPUTR,PRx:PINR

Command/Query

Description

The PRx:PINPUTR command sets the input resistance for both inputs of
the differential amplifier. The valid arguments are 1M or 100M when the
differential amplifier input attenuation is set to ÷1 and an attenuating probe
is not being used. When an attenuating probe is attached to the differential
amplifier, or the internal attenuation is set to ÷10, only 1M may be selected.
The units of the arguments are Ohm.

The PRx:PINPUTR? query returns the input resistance setting for the
differential amplifier connected to the specified channel.

Command Syntax

:PINputR :{PR1,
PR2, PR3, PR4}

:={1M or 100M} when the internal
attenuation is ÷1, without attenuating probe, or ={1M} when the internal
attenuation is ÷10 or an attenuating probe is used.

Query Syntax :PINputR?

Response Format

:PINR

Example

The following command sets the input resistance of the differential
amplifier connected to channel1 to 1 MΩ.

CMD$=”PR1:PINR 1M”:CALL IBWRT(SCOPE%,CMD$)