background image

DA1855A Differential Amplifier

54

922258-00 Rev A

Precision Voltage Generator

PRx:PVGMODE,PRx:PVGM

Command Query

Description

The PRx:PVGMODE command sets the operating mode of the Precision
Voltage Generator (PVG) of the differential amplifier. The valid arguments
are V

COMP

, V

DIFF

and OFF.

The PRx:PVGMODE? query returns the operating mode of the Precision
Voltage Generator (PVG) of the differential amplifier connected to the
specified channel.

Command Syntax

:PVGMode :={PR1, PR2, PR3,
PR4} :={V

COMP

, V

DIFF

or OFF}

Query Syntax

:PVGMode?

Response Format

:PVGM

Example

The following command sets the PVG operating mode of the differential
amplifier connected to channel 3 to V

DIFF

:

CMD$=”PR3:PVGM V

DIFF