beautypg.com

Teledyne LeCroy DA1855A User Manual

Page 77

background image

Operator’s Manual

922258-00 Rev A

71

j. Verify that the channel 1 input coupling is set to DC and 1MΩ.

k. Set the oscilloscope to display channel 1, the vertical sensitivity to 500mV/div,

timebase to 50 ns/div and trigger source to external ÷10. If necessary, adjust the
trigger level for a stable display.

l. Set the leveled sine wave generator frequency to 10 MHz.

m. Set the leveled sine wave generator output amplitude to exactly 2 Vp-p. (4 divisions

on the oscilloscope).

n. Remove the leveled sine wave generator output cable and termination from the

oscilloscope.

o. Attach to the 50 Ω termination a female to female BNC adapter, a BNC ‘Y’ and a 6”

BNC cable to each end of the BNC ‘Y’.

p. Set both the DA1855A +INPUT and –INPUT to DC.

q. Connect the two free ends of the 6” BNC cables to the DA1855A +INPUT and –

INPUT.

r. Reconnect the AMPLIFIER OUTPUT cable to channel 1 of the oscilloscope. Refer to

Figure 19.

Figure 19, HF CMRR Check Setup