beautypg.com

Operator’s manual – Teledyne LeCroy DA1855A User Manual

Page 85

background image

Operator’s Manual

922258-00 Rev A

79

High Frequency CMRR

Step

Description

Intermediate data

Test Result

5-e

Amplifier Output Voltage at 10 MHz

_____________ V

5-f

Differential Mode Gain at
10 MHz

___________

5-w

Common Mode Feedthrough at 10 MHz

___________

5-y

Common Mode Gain at
10 MHz

___________

5-aa

Common Mode Rejection Ratio at 10 MHz (Test limit:

≥ 1000:1)

__________:1

Low Frequency CMRR

6-m

Common Mode Feedthrough at 70 Hz

_____________ V

6-n

Common Mode Gain at
70 Hz

___________

6-o

Common Mode Rejection Ratio at 70 Hz (Test limit:

≥ 50,000:1)

__________:1

6-x

Common Mode Feedthrough at 100 kHz

_____________ V

6-y

Common Mode Gain at
100 kHz

_____________ V

6-z

Common Mode Rejection Ratio at 100 KHz (Test limit

≥ 50,000:1

__________:1

Precision Voltage Generator Accuracy

7-c

PVG Zero Output Voltage

(Test limit: 0.05% of reading + 500 μV)

__________ V

7-g

PVG Output Voltage at +15.5 V

(Test limit: 0.05% of reading + 500 μV)

__________ V

7-j

PVG Output Voltage at

–15.5 V (Test limit: 0.05% of reading + 500 μV)

__________ V