Da1855a test record, Da1855a differential amplifier – Teledyne LeCroy DA1855A User Manual
Page 84

DA1855A Differential Amplifier
78
922258-00 Rev A
DA1855A Test Record
Step
Description
Intermediate data
Test Result
X1 Gain Accuracy
1-f
Sine Wave Generator Output Voltage
_____________ V
1-k
Amplifier Output Voltage
_____________ V
1-m
X1 Gain Error (Test limit
≤ 1%)
_____________%
X10 gain Accuracy
2-g
Sine Wave Generator Output Voltage
_____________ V
2-j
Attenuator Output Voltage
_____________ V
2-l
Exact Attenuation
_____________
2-q
Sine Wave Generator Output Voltage
_____________ V
2-v
Amplifier Input Voltage
_____________ V
2-w
Expected Amplifier Output Voltage
_____________ V
2-x
Measured Amplifier Output Voltage
_____________ V
2-z X10 Gain Error (Test limit
≤ 1%)
_____________%
÷10 Attenuator Accuracy
3-f
Sine Wave Generator Output Voltage
_____________ V
3-g
Expected Output Voltage
_____________ V
3-m
Measured Amplifier Output Voltage
_____________ V
3-o
÷10 Attenuation Error (Test limit
≤± 1.0%)
____________%
X1 Bandwidth and Calculated Rise Time
4-k
Measured -3 dB Frequency at X1 Gain
(Test limit: > 50 MHz)
__________ MHz
4-m
Calculated Rise Time at X1Gain
(Test limit: < 7 nsec)
__________ nsec