Gain Control Mode

The DA1855A amplifiers have two modes for setting the amplifier gain and attenuation; Auto and
Manual. (Refer to Gain Control Modes on page 47, Operation, for more information.) The gain
control mode can be selected in the DA1855A control menu through the oscilloscope front panel, or
through remote control by sending the commands which correspond tot the gain mode.

Sending the PRx:ATTEN or PRx:GAIN commands will set the DA1855A amplifier to Manual gain
control mode.

Sending the PRx:VDIV command will set the DA1855A amplifier to Auto gain control mode.

Attenuation

PRx:ATTENUATION,PRx:ATTEN

Command/Query

Description

The PRx:ATTENUATION command sets the attenuation of the differential
amplifier, including the factor of any passive probe attached to the input.
The command will also switch the Atten/ Gain control mode to Manual if it
was to Auto mode. The valid arguments with no probe attached is 1 or 10
when the input resistance is set 1 MΩ, or only 1 when the input resistance is
set to 100 MΩ. The valid arguments must be scaled by the attenuation
factor of any passive probe which is attached to the amplifier.

The PRx:ATTENUATION? query returns the attenuation of the differential
amplifier, including the attached probe connected to the specified channel.

Command Syntax

:ATTeNuation :={PR1,
PR2, PR3, PR4} :={1,10*}when input resistance is set
to 1 MΩ or {1*} when the input resistance is set to 100 MΩ.

* Attenuation arguments must be scaled by the attenuation of any passive
probe which is attached to the differential amplifier inputs.

Query Syntax

:ATTeNuation?

Response Format

:ATTN

Example

The following command sets the attenuation of the differential amplifier
connected to channel 1 to ÷100 when a ÷10 probe is attached:

CMD$=”PR1:ATTN 100”:CALL IBWRT(SCOPE%,CMD$)