Performance Verification Test Record

This record can be used to record the results of measurements made during the performance
verification of the DA1855A Differential Amplifier. Photocopy this page and record the results on the
copy. File the completed record as required by applicable internal quality procedures.

The section in the test record corresponds to the parameters tested in the performance verification
procedure. The numbers preceding the individual data records correspond to the steps in the
procedure that require the recording of data. Results to be recorded in the column labeled "Test
Result" are the actual specification limit check. The test limits are included in all of these steps.

Other measurements and the results of intermediate calculations that support the limit check are to
be recorded in the column labeled "Intermediate Results".

Permission is granted to reproduce these pages for the purpose of recording test results.

Serial Number:

DA1855A

Asset or Tracking Number:

Date:

Technician:

Equipment Used

MODEL

SERIAL NUMBER

CALIBRATION

DUE DATE

OSCILLOSCOPE

PREAMPLIFIER

DIGITAL MULTIMETER

LEVELED SINE WAVE
GENERATOR

HIGH AMPLITUDE SINE
WAVE GENERATOR 1

N/A

1 The high amplitude sine wave generator is used in this Performance Verification Procedure for making relative
measurements. The output of this generator is measured with a DMM or oscilloscope in this procedure. Thus, the
generator is not required to be calibrated.