Operator’s Manual

922258-00 Rev A

5

General, continued

Auto Zero Amplifier initiates an automatic balance cycle, when either gain

button is depressed, to remove output offset drift

Effective Gain Indicator Indicators show the effective system gain or attenuation, factoring

Probe Attenuation, Attenuator and gain settings. (Probe must have
coding connectors. ч1, ч10, ч100 and ÷1000 probes are
recognized)

Dynamic Ranges

Maximum Differential Mode Range

X10 Gain, ÷1 Attenuator ± 50 mV1

X1 Gain, ÷1 Attenuator ± 0.5 V1

X10 Gain, ÷10 Attenuator ± 0.5 V1

X1 Gain, ÷10 Attenuator ± 5 V1

Maximum Input Slew Rate

÷1 Attenuator

± 0.15 V/μsec

÷10 Attenuator

± 1.5 V/μsec

Maximum Common Mode Range

÷1 Attenuator ± 15.5 V

1

÷10 Attenuator

± 155 V

1

Differential Offset Range (V

DIFF

mode referred to input)

X10 Gain, ÷1 Attenuator ± 1 V

1

X1 Gain, ÷1 Attenuator ± 10 V

1

X10 Gain, ÷10 Attenuator ± 10 V

1

X1 Gain, ÷10 Attenuator ± 100 V

1

Comparison Offset Range (V

COMP

mode, referred to input)

÷1 Attenuator ± 15.5 V

1

÷10 Attenuator ± 155 V

1

1 Voltages are referred to the amplifier input connector. Multiply by probe attenuation factor to obtain value referred to
probe input (e.g. ± 50 mV becomes ± 0.5 V at the probe tip when using a ÷10probe.)