Micromod Micro-DCI: 53ML5100A LOADING STATION User Manual
Page 38
When the HHC-to-controller transfer begins a message is displayed indicating the transfer
has started. Another message appears when the transfer is completed. If the transfer
fails, an error message is NOT displayed to indicate the transfer failed; the transfer com-
pleted message is not displayed after 12 seconds (approximate). If the cause of the trans-
fer error is corrupted data on the HHC cartridge, then replace it with a backup copy and
proceed with the transfer sequence again; otherwise, there may be a problem with the
HHC, the connecting cable, or instrument RS232 configuration port.
2.
If F2 is selected, the following menu appears:
F2-DATABASE
F3-PROGRAM (not applicable)
Based on which selection from this menu is made the following operations can occur:
•
F2- DATABASE: This operation copies the entire database, all B, L, C, H, and A, data-
point values, in the instrument to the storage cartridge. Transfer takes approximately 12
seconds.
•
F3- PROGRAM: This operation is not applicable for this instrument.
When the controller-to-HHC transfer begins a message is displayed indicating the transfer
has started. Another message appears when the transfer is completed. If the transfer
fails, an error message is displayed instead of the transfer completed message. If an error
does occur, resolve the cause and re-initiate the transfer. A transfer error should not be ig-
nored, as a partial transfer may cause the HHC cartridge to contain corrupted data. The
HHC returns to the command entry mode if the transfer is successful or if it is halted with
an error message.
Section 3. Displays and Push Buttons
3-15