Table 1-1. specifications – Micromod Micro-DCI: 53ML5100A LOADING STATION User Manual
Page 14
Table 1-1. Specifications
Item
Specification(s)
Environmental Characteristics
Controlled Environment
Ambient Temperature Limits
Relative Humidity Limits
Temperature Effects on Accuracy
Transient Immunity (all circuits)
EMI Susceptibility
Enclosure Classification/Environment
Shock
Vibration
Drop and Topple
Enclosed temperature controlled location (Class
A and B per ISA-S71.01 1985)
4 - 52
°
C (40 - 125
°
F)
10 - 90% maximum
±
0.28% per 28
°
C (50
°
F) change from reference
temperature 25
°
C (77
°
F)
ANSI C37.90a - 1974/IEEE Std 472-974:
Ring Wave: 1.5 MHz, 3 kV, 60 pulses/s for 2.0 s
SAMA PMC 33.1-1978: Class 3-abc: no effect
at 30 V/m, at 27, 146, and 446 MHz
Panel Mounted Equipment: No enclosure
rating. Designed to be installed in a user
provided panel or enclosure.
Rated for installation in a Pollution Degree 2
location per U.L. 508-1989/Controlled
Environment per CSA C22.2 No. 142-M1987.
An indoor, temperature controlled location
(Control Room or Shop Floor) where normally
only non-conductive pollution occurs; however,
temporary conductivity caused by condensation
may be expected.
Location in environments more severe than
those stated requires supplementary protection.
0.5 g
SAMA PMC 31.1-1978: peak-to-peak constant
displacement 0.05 in. (1.27 mm), 5-14 Hz:
0.5 g, 14-200 Hz.
SAMA PMC 31.1-1978: Tilt 30 degrees from
horizontal and fall freely to a hard surface, all
sides, front, and back.
53ML5100 Manual Loading Station