Storing the luminex® flexmap 3d® instrument, Storing the luminex, Flexmap 3d – Luminex FLEXMAP 3D User Manual User Manual
Page 48: Instrument
![background image](/manuals/336890/48/background.png)
6. Plug in the power cord and turn on the FLEXMAP 3D
®
instrument.
7. Use the software to run the prime command twice.
FIGURE 20.
Front of Fluidics Bay (Inside Left Door)
1
Top quick-disconnect point
2
Sheath Filter
3
Bottom quick-disconnect point
Storing the Luminex
®
FLEXMAP 3D
®
Instrument
If you need to put the Luminex
®
FLEXMAP 3D
®
instrument in long-term storage or prepare it
for use after removing it from long-term storage, use the following procedures.
Storing the Luminex
®
FLEXMAP 3D
®
Instrument
This procedure explains the steps you should take before placing the Luminex
®
FLEXMAP
3D
®
instrument into long-term storage.
To prepare the FLEXMAP 3D
®
instrument for storage:
1. Use the software to perform a Preparation for storage routine.
Luminex
®
FLEXMAP 3D
®
Hardware User Manual
38
For research use only. Not for use
in diagnostic procedures.