Storing the luminex® flexmap 3d® instrument, Storing the luminex, Flexmap 3d – Luminex FLEXMAP 3D User Manual User Manual

Page 48: Instrument

background image

6. Plug in the power cord and turn on the FLEXMAP 3D

®

instrument.

7. Use the software to run the prime command twice.

FIGURE 20.

Front of Fluidics Bay (Inside Left Door)

1

Top quick-disconnect point

2

Sheath Filter

3

Bottom quick-disconnect point

Storing the Luminex

®

FLEXMAP 3D

®

Instrument

If you need to put the Luminex

®

FLEXMAP 3D

®

instrument in long-term storage or prepare it

for use after removing it from long-term storage, use the following procedures.

Storing the Luminex

®

FLEXMAP 3D

®

Instrument

This procedure explains the steps you should take before placing the Luminex

®

FLEXMAP

3D

®

instrument into long-term storage.

To prepare the FLEXMAP 3D

®

instrument for storage:

1. Use the software to perform a Preparation for storage routine.

Luminex

®

FLEXMAP 3D

®

Hardware User Manual

38

For research use only. Not for use

in diagnostic procedures.