Initializing the luminex® flexmap 3d® instrument, Warming up the luminex® flexmap 3d® instrument, Maintaining fluids – Luminex FLEXMAP 3D User Manual User Manual
Page 38: Initializing the luminex, Flexmap 3d, Instrument, Warming up the luminex

the performance of these activities, see the appropriate Luminex
software manual or Luminex online help.
Initializing the Luminex
®
FLEXMAP 3D
®
Instrument
NOTE: Use the software to perform system initialization.
The software provides three options for system initialization:
• Warmup, fluidics
• Warmup, fluidics, verification
• Warmup, fluidics, verification, and calibration
Select the option you want. Warmup, fluidics, and verification should be part of the daily
maintenance of the instrument; calibration can be limited to weekly maintenance unless
otherwise specified by assay instructions.
NOTE: If the instrument temperature falls out of the delta cal range,
recalibrate it. For information about the delta cal range, see the
appropriate Luminex software manual or Luminex online help.
Warming Up the Luminex
®
FLEXMAP 3D
®
Instrument
NOTE: If the Luminex
®
FLEXMAP 3D
®
instrument is powered on but idle for
more than four hours, both lasers turn off. Consequently, the
instrument should be warmed up again. Use the software to perform
a warmup.
Maintaining Fluids
Periodically monitor fluid levels. Replace the empty sheath fluid container as needed. If the
Luminex
®
FLEXMAP 3D
®
instrument operates with an empty sheath fluid container, the lack
of sheath fluid may interrupt a sample and prevent further samples from being collected.
CAUTION:
Use only xMAP Sheath Fluid or another Luminex-approved
sheath fluid. Use of any other sheath fluid constitutes improper
use and can void the warranty provided by Luminex and/or its
authorized partner.
Optimally, the waste line should be routed to a laboratory drain. If you use a waste container,
monitor waste fluid levels periodically. The volume of the waste container should be at least
as large as the volume of the sheath container. Do not allow the waste fluid container to
overflow. Empty the waste fluid container each time you replace or fill the sheath fluid
container.
Place the waste fluid container at least three feet below the surface on which the FLEXMAP
3D
®
instrument rests.
WARNING:
Do not place the waste fluid container on top of the instrument.
Luminex
®
FLEXMAP 3D
®
Hardware User Manual
28
For research use only. Not for use
in diagnostic procedures.