Replacing the HEPA Air Filter

The HEPA air filter is the round filter behind the panel on the left side of the front of the
fluidics bay. Tubing from the interior of the fluidics bay is attached to the stem in the center
back of the filter.

To replace the HEPA air filter:

1. Turn off the Luminex

®

FLEXMAP 3D

®

instrument and unplug the power cord.

2. Open the left door and locate the HEPA air filter panel on the front of the fluidics bay as

shown in the HEPA Air Filter image below.

3. Remove the screw at the top of the panel and open the panel door as shown in the

HEPA Air Filter image below.

4. Grasp the tubing and pull the filter 3 to 4 inches from the unit.

5. Remove the filter with one hand and hold the tubing with the other hand.

CAUTION:

Do not allow the tubing to fall inside the instrument.

6. Connect a new filter to the tubing and position the filter inside the panel.

7. Reattach the panel door to the unit and close the access door.

8. Plug in the power cord and turn on the FLEXMAP 3D

®

instrument.

FIGURE 17.

HEPA Air Filter

View A

Front of fluidics bay

View B

HEPA Filter

1

Filter panel

3

Filter Stem

2

Screw - Remove as indicated in Step 3.

4

Attachment point for tubing

Luminex

®

FLEXMAP 3D

®

Hardware User Manual

34