Printer, Barcode labels, Vortex – Luminex FLEXMAP 3D User Manual User Manual
Page 32: Bath sonicator, Specifications and limitations, General, Printer barcode labels vortex bath sonicator

Printer
Use a printer compatible with Microsoft
®
Windows XP SP3.
Barcode Labels
Use Code 128 barcode label type when scanning barcode labels into the system.
Vortex
Use VWR product number 58816-12, with a speed range of 0 to 3200 rpm, or equivalent.
Bath Sonicator
Use Cole-Parmer
®
product number 08849-00, with an operating frequency of 55 kHz, or
equivalent.
Specifications and Limitations
The Luminex
®
FLEXMAP 3D
®
system is designed to fit a series of standards for speed,
accuracy, sensitivity, and capacity. The following topics provide the specifications and
limitations for the Luminex FLEXMAP 3D
®
system.
CAUTION:
Read and observe specifications and limitations carefully.
General
Indoor laboratory and professional use only
Physical dimensions:
58.4 cm (23 inches) W x 65.3 cm (25.7 inches deep, including the required 1.25 inches for
proper cooling) D x 54.7 cm (18.0 inches) H. Additional space required for the arm, monitor,
keyboard, mouse and optional barcode scanner does not exceed 64.8 cm (25.5 inches) W
by 61 cm (24.0 inches) D.
Weight:
up to 91 kg (200 lbs), including the instrument, monitor arm, monitor and accessories such
as the keyboard, mouse, and barcode scanner.
Installation Category:
II
Pollution degree:
2
Operating temperature:
15°C to 30°C (59°F to 86°F)
Operating humidity:
20% to 80%, noncondensing
Shipping and storage
temperature:
0 to 50°C (32° F to 122° F)
Shipping and storage
humidity:
20% to 80%, noncondensing
Luminex
®
FLEXMAP 3D
®
Hardware User Manual
22
For research use only. Not for use
in diagnostic procedures.