beautypg.com

Printer, Barcode labels, Vortex – Luminex FLEXMAP 3D User Manual User Manual

Page 32: Bath sonicator, Specifications and limitations, General, Printer barcode labels vortex bath sonicator

background image

Printer

Use a printer compatible with Microsoft

®

Windows XP SP3.

Barcode Labels

Use Code 128 barcode label type when scanning barcode labels into the system.

Vortex

Use VWR product number 58816-12, with a speed range of 0 to 3200 rpm, or equivalent.

Bath Sonicator

Use Cole-Parmer

®

product number 08849-00, with an operating frequency of 55 kHz, or

equivalent.

Specifications and Limitations

The Luminex

®

FLEXMAP 3D

®

system is designed to fit a series of standards for speed,

accuracy, sensitivity, and capacity. The following topics provide the specifications and
limitations for the Luminex FLEXMAP 3D

®

system.

CAUTION:

Read and observe specifications and limitations carefully.

General

Indoor laboratory and professional use only

Physical dimensions:

58.4 cm (23 inches) W x 65.3 cm (25.7 inches deep, including the required 1.25 inches for
proper cooling) D x 54.7 cm (18.0 inches) H. Additional space required for the arm, monitor,
keyboard, mouse and optional barcode scanner does not exceed 64.8 cm (25.5 inches) W
by 61 cm (24.0 inches) D.

Weight:

up to 91 kg (200 lbs), including the instrument, monitor arm, monitor and accessories such
as the keyboard, mouse, and barcode scanner.

Installation Category:

II

Pollution degree:

2

Operating temperature:

15°C to 30°C (59°F to 86°F)

Operating humidity:

20% to 80%, noncondensing

Shipping and storage
temperature:

0 to 50°C (32° F to 122° F)

Shipping and storage
humidity:

20% to 80%, noncondensing

Luminex

®

FLEXMAP 3D

®

Hardware User Manual

22

For research use only. Not for use

in diagnostic procedures.