3 sample factory calibration sheet – Liquid Controls Sponsler T675 - Cryogenic System Register User Manual
Page 43

Sponsler, Inc.
Model T675 Flow Totalizer
Technical Information
Pg. 40
5.3 SAMPLE FACTORY CALIBRATION SHEET
WORK ORDER #
Sponsler, Inc.
LUCITE
A Unit of IDEX Corporation
Calibration :
2 Weight(Lbs):
200.0 Serial # :
KENDEMO
Fluid :
WATER Volume Sp. Gr. 1.000
Housing Material :
LUCITE
Test Stand # :
3 (Gallons Water):
23.9926103 Rotor Material :
NICKEL
Calib. By :
SWR
Model #
Bearings :
CRYO
Calib. Date :
11/20/97
SP2-
Coil # :
PC13-105G (1EA.)
PT.
(TRUE)
TEMP.
TRUE VOL.
RATE
TIME
TOTAL
APPROX.
"K"
SP.GR.
DEG.FAR.
GAL
GPM
SECONDS
PULSES
FREQ.HZ.
GAL
1
0.9983
67.1 24.03279 227.04575
6.351
3576 563.06094 148.79669
2
0.9984
66.4 24.03092 202.28043
7.128
3585 502.94613 149.18283
3
0.9985
65.6 24.02863 179.05089
8.052
3590 445.85196 149.40511
4
0.9985
65.7 24.02892 155.39287
9.278
3587 386.61349 149.27847
5
0.9985
65.7 24.02911 132.56221
10.876
3574 328.61346 148.73626
6
0.9985
65.9 24.02945 108.24888
13.319
3574 268.33846 148.73417
7
0.9985
65.9 24.02962
85.81494
16.801
3568 212.36830 148.48344
8
0.9984
66.1 24.03000
61.34015
23.505
3586 152.56329 149.23012
9
0.9984
66.1 24.03015
38.85334
37.109
3567 96.12223 148.43855
10
0.9984
66.2 24.03036
15.21471
94.765
3577 37.74600 148.85335
MEAN TOTAL PULSES (PER WEIGHT USED):
3578.40
SENS.I.NG ELEMENT CONSTANT (MEAN "K"):
148.91390
Remarks:
Conversion Details: GALLONS (US)
Range Bars: +/-0.50%
Conversion Medium: Water
The above mentioned item has been calibrated with equipment which has an uncertainty of +/- 0.11% and is traceable to the National Institute of
Standards and Tech. per MIl-STD-45662A through test numbers SC721201-22, 30, 33, 34, 37-39M; SC721122-17M; C7881120-4M; SC860617-1M;
SC730808-7M; SC720107-25M; SC721101-13M, 31M, 54M, 57M; SC730322-3M; AND SC730122-14M, 18M. Also through NIST certificate numbers
731/243669; 807675; 86707942; 711/254789; 254366, 67; 251971; 253652; and WWV. Specific gravity referenced at 39.2 degrees Fahrenheit.
* Runs not included in the mean pulses or sensing element constant. # Interpolated runs. X Run outside meter accuracy.
CALIBRATED BY:
SCI Equipment Calibration:
CERTIFIED BY:
Last:
Oct-97
DATE:
03/21/06
Next:
Apr-98