beautypg.com

Default settings, Menu structure – Hach-Lange EVITA OXY Basic User Manual User Manual

Page 98

background image

English 98

Default settings

The EVITA® OXY system is supplied with the following default settings.

Menu structure

The menu of the signal converter is divided into two parts. The operator
menu and the setup menu.

Operator menu

The operator menu is designed for daily use. After the language has
been selected, the signal converter begins each time in the operator
menu and shows the actual concentration of dissolved oxygen.
The model USC 5000 is designed for point-to-point installation, the
model USC 7000 for multidrop systems. In

Figure 19, page 99

, gray

marked menus are only visible via the multidrop system. There is also
the option to hide some menus within the operator menu (refer to

Figure 29, page 108

).

USC 5000/7000

Parameter

Default
setting

Options

Current output 1
(System package 2)

0–20 mg/L

Range:

Measure
ment
range:

minimum 0 mg/L or ppm;
0 %
maximum 50 mg/L or ppm;
500 %
minimum 1 mg/L or ppm;
10 %

Current output 2

0–40 °C

Range:

Measure
ment
range:

minimum -10 °C
maximum 70 °C
minimum 1 °C

Current outputs
USC 5000/7000

OFF

OFF
ON

Relay 1, 2, 3, 4 and 5
(USC 7000)

OFF

Error, warning, limit, timer, OFF

TILTCAL®

Activated Activated;

Deactivated

Passcode 1000

1000–9999

OXY 4100/4150

Parameter

Default
setting

Options

Oxygen measurement
unit

mg/L

mg/L; ppm; %

Temperature
measurement unit

°C

°C; °F

Current output, if no
measurement

Low

High; Low; Hold
(last measurement value)

Current output if fault

Low

High; Low; Normal

Time constant

40 seconds

10–300 seconds