Usc 7000 multidrop system, English 102 – Hach-Lange EVITA OXY Basic User Manual User Manual
Page 102

English 102
USC 7000 multidrop system
Change transmitter address
So that the USC 7000 can detect the transmitter via the HART
®
bus, a
clear address value between 1 and 15 must be assigned to the
transmitter. The transmitter EVITA
®
OXY is supplied with the default
setting 0, which is used for point-to-point installations (USC 5000).
For the programming of the address of a new transmitter connect this to
the USC 7000 and select with the keys of the USC the menu item "Scan
HART bus" (
.) The next available address between
1 and 15 is assigned to the new transmitter.
Note:
If before the access to the HART
®
bus an OXY transmitter with an
address value that is over 0 has been registered as interrupted for longer
than two minutes, this address is assigned to the new OXY transmitter.
Figure 21 Transmitter address
Setup of the oxygen- and temperature units and of the
current output of the transmitter
If required the default settings can be changed as shown below.
The setup of the display- and current output for oxygen- and
temperature measurement device is done via "Basic Settings". The
setup of the transmitter measurement range for the oxygen content is
done via change of the values 4 mA and 20 mA.
Otherwise the setup of the measurement range of the oxygen- and
temperature measurement unit is done as shown in
and in
Figure 22 Transmitter settings
Rem. lifetime
Transmitter 0
72. h
No. of alarms
Transmitter 0
12. d
Calibration
Transmitter 0
42. d
Temperature
Transmitter 0
Oxygen
Transmitter 0
10.64 mg/l
Error pending
Scan HART bus
Password
0000.
Settings
USC
Basic settings
TR. 0
Advanced settings
TR. 0
Settings OXY
Ser.No. SSSSVVNUUÅ
TR. 0
Set transmitter
address 00
TR. 0
081R9302.02.02
Reset lifetime
Transmitter 0
Last calibration
Transmitter 0
USC
Scan HART bus
Are you sure ?
USC
Scan HART bus
Busy
USC
Trm. address 0
changed to 05