Hach-Lange EVITA OXY Basic User Manual User Manual
Page 118

English 118
Flashing bell symbol
in the display
Not defined
WU10
I OUT 1: Values at 4 mA \>= values at 20 mA
Check setup of I OUT 1
WU11
I OUT 2: Values at 4 mA \>= values at 20 mA
Check setup of I OUT 2
WU12
I OUT 3: Values at 4 mA \>= values at 20 mA
Check setup of I OUT 3
WU13
I OUT 4: Values at 4 mA \>= values at 20 mA
Check setup of I OUT 4
WU14
Measurement value above current output setting 1 at 20 mA
Change values at 20 mA for I OUT 1
WU15
Measurement value above current output setting 2 at 20 mA
Change values at 20 mA for I OUT 2
WU16
Measurement value above current output setting 3 at 20 mA
Change values at 20 mA for I OUT 3
WU17
Measurement value above current output setting 4 at 20 mA
Change values at 20 mA for I OUT 4
WU18
Measurement value above current output setting 1 at 4 mA
Change values at 4 mA for I OUT 1
WU19
Measurement value above current output setting 2 at 4 mA
Change values at 4 mA for I OUT 2
WU20
Measurement value above current output setting 3 at 4 mA
Change values at 4 mA for I OUT 3
WU21
Measurement value above current output setting 4 at 4 mA
Change values at 4 mA for I OUT 4
WU39
HART data exchange error
Check cabling
WU40
TMS data exchange error
Replace TMS
WU41
A parameter is outside of the range.
Could not be replaced by the standard value.
Is remedied by new switch-on
WU43
Too many errors simultaneously.
Some errors are not correctly registered.
Switch USC off and on again
WU49
Internal error
Switch USC off and on again
One or several
outputs = 3.8 mA
or 24 mA
EU60
The CAN bus application is not responding.
This error is shown when the maximum number of
communication attempts has been exceeded.
1. Switch USC off and on again
2. Replace USC
Symptom
Current output
Event
code
Cause
Resolution