Flashing bell symbol
in the display
Not defined
EU61
USC cannot read the data for current output
I OUT 1
1
No response from the transmitter
2
Transmitter has received a new address
3
Another master is calling up problems with the
HART
®
communication.
4
Supply voltage for transmitter
too low
5
Too much noise interference on the current loop
6
Transmitter defective
7
USC defective
1
Check cable and shielding
2
Switch USC off and on again
3
Disconnect other master, or set up
USC for other master, refer to figure
51, page 52
4
Check supply voltage of the transmit-
ter; it must be at least 20 V.
5
Check cable and shielding
6
Replace transmitter
7
Replace USC
EU62
USC cannot read the data for current output I OUT 2
1–7: As with EU61
EU63
USC cannot read the data for current output I OUT 3
1–7: As with EU61
EU64
USC cannot read the data for current output I OUT 4
1–7: As with EU61
EU65
USC cannot read the data for relay 1
As with EU61
EU66
USC cannot read the data for relay 2
As with EU61
EU67
USC cannot read the data for relay 3
As with EU61
EU68
USC cannot read the data for relay 4
As with EU61
EU69
USC cannot read the data for relay 5
As with EU61
Symptom
Current output
Event
code
Cause
Resolution