Usc 7000 relay output settings, English 105 – Hach-Lange EVITA OXY Basic User Manual User Manual
Page 105

English 105
USC 7000 relay output settings
Relay output settings 1 to 5 for USC 7000.
Figure 25 Relay output USC 7000
* Units selected in the section "Default settings of the transmitter" (
** So that the current output can stabilize after a cleaning, the USC will define the current output during "cleaning time" and another 10 minutes.
Output
USC
I out 1
USC
Timer interval
000. h
RELAY 1
1-999 hours
Active time
00 min
RELAY 1
1-59 min
Relay at
active Open
RELAY 1
Open/Closed
Relay at
warning Open
Open/Closed
Function
Error
RELAY 1
RELAY 1
Relay at
error Open
RELAY 1
Open/Closed
RELAY 1
Setpoint
050.00 mg/l
RELAY 1
0 - 999.99
Hysteresis
005.00 mg/l
RELAY 1
0 - 999.99
Relay above
limit Open
RELAY 1
Open/Closed
Relay 1
USC
Relay 2
USC
Relay 3
USC
~~
Relay below
limit Open
RELAY 1
Open/Closed
081R9415.02.02
Relay 4
USC
Relay 5
USC
Function
Off
RELAY 1
One transmitter
All Errors
All Warnings
Timer
Off
Relay at
warning Open
Open/Closed
RELAY 1
Relay at
Error Open
RELAY 1
Open/Closed
Select
transmitter no. 00
RELAY 1
00-15
Error
Warning
Hi Limit
Lo Limit
Cleaning
Oxygen
Temperature
Value
Oxygen
Setpoint
035.00 ºC
RELAY 1
Hysteresis
005.00 ºC
RELAY 1
0 - 999.99
0 - 999.99
*
*
*
*
Cleaning interval
000. h
RELAY 1
Cleaning time
00 min
RELAY 1
Relay at
cleaning Open
RELAY 1
1-999 h
1-59 min
Open/Closed
**