7

Signet 0250 Setup Tool

Signet 2450 Pressure Sensor Setup

Windows PC

Signet 2450

Pressure

Sensor

USB

GND

Data

+5V out

GND

Data

+5V

2450 Wiring

Operation
1. Select the engineering units for the application: kpa, bar

or psi

2. Set the Pressure measurement where the Loop output

must equal 4 mA.
• Minimum setting 0 psi
• Maximum setting 10, 50 or 250 psi

3. Set the Pressure measurement where the Loop output

must equal 20 mA.
• Minimum setting 0 psi
• Maximum setting 10, 50 or 250 psi

4. Click "Write" to copy the settings from the display into the

Magmeter.
• Click "Save" to save the settings from the display to a

computer fi le.

To use a saved fi le:
• Click "Load"
• Navigate to the saved fi le
• Select Open
• Click "Write"
• Click "Read" to confi rm

Signet 2750 pH Sensor Setup

2750 Wiring

2750 Operation
1. Select the engineering units for the application: pH or ORP
2. Set the pH or ORP measurement where the Loop output

must equal 4 mA.
• Minimum setting 0 pH, -1000 mV ORP
• Maximum setting 14 pH, 2000 mV ORP

3. Set the pH or ORP measurement where the Loop output

must equal 20 mA.
• Minimum setting 0 pH, -1000 mV ORP
• Maximum setting 14 pH, 2000 mV ORP

4. Click "Write" to copy the settings from the display into the

Magmeter.
• Press "Save" to save the settings from the display to a fi le.

To use a saved fi le:
• Click "Load"
• Navigate to the saved fi le
• Select Open
• Click "Write"
• Click "Read" to confi rm.